Summary of Contents for nt-023

Page 1: ......

Page 2: ...MENT 7 5 1 Distance 7 5 2 Stake Out S O 7 5 3 Mode 8 6 AXES STAKE OUT 9 6 1 Stake Out of One Side on the Axes 9 6 2 Stake Out of Any Point 11 7 QUICK SETTING 13 7 1 Laser Plummet 13 7 2 Laser Pointer...

Page 3: ...6 9 CALIBRATION 17 9 1 Calibrate i Angle 17 9 2 Correction of Additive Constant 18 10 INFORMATION 20 10 1 Firmware Upgrade 20 10 2 Factory Mode 21 10 3 System Information 21 11 SPECIFICATION 22 12 ERR...

Page 4: ...adapts to the temperature of environment e If the instrument has not been used for a long time you should remove the battery for separate storage The battery should be charged once a month f When tra...

Page 5: ...ment abnormal l Do not aim the laser beam to eyes m Keep the screen clean Do not scratch the screen with sharp objects n This products belongs to the Level II laser products It fully meet the requirme...

Page 6: ...PART NAMES 1 Collimator 2 Objective Lens 3 Focusing Screw 4 Telescope Eyepiece 5 Horizontal Tangent Screw 6 Battery 7 RS232 Interface 8 Display Unit 9 Tribrach 10 Tribrach Lock 11 Vertical Tangent Sc...

Page 7: ...hortcut to set the laser plummet 2 Number 2 Shortcut to set the laser pointer 3 Number 3 Shortcut to set the compensation on Axis X N 4 Number 4 Shortcut to set distance measurement mode 5 Number 5 Sh...

Page 8: ...ance HD horizontal distance SD slope distance hPa unit of air pressure hectopascal mmHg unit of air pressure millimeter of mercury inHg unit of air pressure inch of mercury m unit of distance meter ft...

Page 9: ...slope V switching Face Right and Face Left R L etc 4 1 Angle 0Set Set the current angle to 0 HSet Input an angle to set as the current horizontal angle V degree or gradient display R L Shift Face Lef...

Page 10: ...etc 5 1 Distance Aim at the center of the target prism through the optical eyepiece by adjusting the focus and press Meas to start the distance measurement Meas Start to measure the distance S O Start...

Page 11: ...Save the input value and continue to stake out Display of Stake Out Result VD The distance difference between the current horizontal distance and the horizontal distance about to stake out 5 3 Mode Th...

Page 12: ...re 2 options to define the baseline One is to define by station point and a known bearing angle 0 00 00 one is to define by two new points Press F1 or F2 to select 6 1 Stake Out of One Side on the Ax...

Page 13: ...long the axes from Point A to B Offst Offset value perpendicular to the axis Step 4 Press F1 to measure Move the prism according to the indications on the screen until all the values on the screen are...

Page 14: ...along the axis SwPt Return to input offset to start a new point 6 2 Stake Out of Any Point Stake out in this mode can set the theodolite at any point outside the axis Select F2 Step 1 Measure the dis...

Page 15: ...2 Input the line and offset values press F4 to next step Step 3 Press F1 to measure Indicate the poleman to move the prism according to the indications on the screen until all the values on the scree...

Page 16: ...et to enter this page There re 5 settings in QuickSet laser plummet laser pointer compensation on X axes distance setting backlight and sound 7 1 Laser Plummet Press F1 for laser plummet setting You c...

Page 17: ...n X axis Turn on and off the compensation on X axis and check the tilt value 7 4 Distance Setting Set the prism constant measure mode and measure times under the distance setting page 7 5 Backlight an...

Page 18: ...Press Menu then F2 to enter the SET page It covers 5 settings unit angle distance PPM and power set 8 1 Unit Set the units of angle distance temperature and air pressure 8 2 Angle Set the display of v...

Page 19: ...e factor Ht Elevation of the station point 8 4 PPM Set parameters related to temperature and air pressure 8 5 Power Set the sleep time off time Check the battery capacity directly SlepTime Time to ent...

Page 20: ...vertical index difference Step 1 On Face Left collimate the crosshair center in a collimator and adjust the focus until it is clear Press OK to proceed to next step Step 2 Turn the theodolite to Face...

Page 21: ...specting it once or twice a year You can follow the following steps to do a quick inspection and correction Inspection Step 1 Set the instrument on a flat ground mark it as Point A Along the vertical...

Page 22: ...K should be close to 0 If k 5mm the theodolite should be delivered to professional workshop which has standard alignment to calibrate Correction According to the formula input the K value in the foll...

Page 23: ...are and check the system information 10 1 Firmware Upgrade Firmware upgrade includes angle system and distance system Press F1 to enter the firmware upgrade page It is required to connect the theodoli...

Page 24: ...upgrading software 10 2 Factory Mode Press F2 to enter the factory mode and reset to default setting Input the password to proceed 10 3 System Information Press F3 to check the system information The...

Page 25: ...R Wave Length 635 20nm Class II Diameter of Laser Dot 5mm 100m Accuracy 10 ANGLE Type Absolute Encoding Diameter of Disk 79mm Detecting Method Horizontal dual Vertical dual DISTANCE Reflecting Target...

Page 26: ...LUMMET Type Class II visible red laser Wave of Length 635 20nm Accuracy 1 5mm when HT 1 5m Diameter 2 5mm when HT 1 5m DISPLAY Type 160 96 Dot Matrix Size 2 7 inch POWER Type Li on Voltage 7 4V Workin...

Page 27: ...r H CCD Error V Rotate ERR H Rotate ERR Distance Problem ERR32 Dimmer motor error ERR33 High voltage error ERR35 Internal light path error ERR38 Internal light path error ERR41 Temperature sensor erro...

Page 28: ......