background image

________________________________________________________________ 

Medix T22B & T24B

Battery Contact PCB Replacement 

Instructions 

May 2017 

Summary of Contents for Medix T22B

Page 1: ...________________________________________________________________ Medix T22B T24B Battery Contact PCB Replacement Instructions May 2017...

Page 2: ...instructions may not be exactly the same as the product you have but the procedures should be the same NOTE PREPARE A CLEAN WORKING SURFACE TO PROTECT SYSTEM S DISPLAY CASE FROM SCRATCHES DAMAGE If a...

Page 3: ...Front Bezel is attached to the system by snap clips Put fingers on each side of the Front Bezel and gently pull it up Turn the system over with the screen facing down 4 Remove a M3 x 10 mm screw Remo...

Page 4: ...ct PCB Replacement Instructions 4 5 Remove 2 M3 x 10 mm screws Remove the HDD cover If there is no HDD installed you do not need to remove the HDD cover If there is the HDD in the compartment disengag...

Page 5: ...Medix T2xB Series Battery Contact PCB Replacement Instructions 5 6 Remove the base cover a Remove 11 M3 x 10 mm screws for Medix T22B Remove 15 M3 x 10 mm screws for Medix T24B...

Page 6: ...xB Series Battery Contact PCB Replacement Instructions 6 b Remove 3 M3 x 10 mm screws for Medix T22B Remove 3 M3 x 10 mm screws and 2 M3 x 6 mm screws for Medix T24B c Remove base cover from the syste...

Page 7: ...ttery Contact PCB Replacement Instructions 7 7 For replacing the Battery Contact PCB a Remove 6 M3 x 6 mm screws 2 screws for each PCB Disconnect the cables by gently dislodging the connectors from th...

Page 8: ...T2xB SeriesBattery Contact PCB Replacement Instructions 8 b Place new Battery Contact PCBs into the bracket Apply 6 screws and engage the cables For re assembly follow the previous instructions in re...

Reviews: