background image

 

 

 

HH-521BT 

4-Channel Thermocouple Data Logger (with Bluetooth) 

(Types K, J, T, N, E, R, S) 

Instruction Manual 

 

 

Summary of Contents for HH-521BT

Page 1: ...HH 521BT 4 Channel Thermocouple Data Logger with Bluetooth Types K J T N E R S Instruction Manual...

Page 2: ...utton 7 7 OPERATING INSTRUCTIONS 7 7 1 Setup Options 7 7 2 Menu Item 7 7 3 Menu Description 9 7 3 1 Set Bluetooth ON OFF 9 7 3 2 Select thermocouple type K J E T N R or S 9 7 3 3 Setting interval time...

Page 3: ...ction provided by the meter may be impaired ENVIRONMENT CONDITIONS Altitude up to 2000 meters Relatively humidity 90 max Operation ambient temperature 0 C to 50 C MAINTENANCE CLEANING Repairs or servi...

Page 4: ...F 0 5 of reading 0 7 C 0 5 of reading 1 3 F R and S type 0 2 of reading 1 4 C 0 2 of reading 2 5 F Temperature coefficient 0 01 of reading 0 05 C 0 0028 F per C 18 C 64 F or 28 C 82 F Sample rate 2 ti...

Page 5: ...ogger indicator Memory full indicator Memory group indicator Recall Memory group reading indicator Thermocouple type Temperature reading Temperature channel Temperature subtraction mode Temperature al...

Page 6: ...Thermocouple Input 9 RECALL Saved Reading Button 2 Display Screen 10 SETUP Button 3 Power ON OFF Button 11 MEM Button Set 100 Memory 4 Back Light Button 12 USB Interface 5 MAX MIN AVG Button 13 Blueto...

Page 7: ...e During recording period most of the buttons are disabled such as the All settings must be made before starting the data logger function Note When the memory is full 32 000 recorders FULL symbol will...

Page 8: ...ess button again the MAX MIN and AVG symbol blinks together the readings shows real time data To exit MAX MIN AVG mode Press and hold button for 2 seconds to exit MAX MIN AVG mode 6 8 Selecting Temper...

Page 9: ...uction Manual v1 1 Thermosense Limited www thermosense co uk Page 8 2 Fig 2 Set thermocouple type 3 Fig 3 Set interval time for data logging 4 Fig 4 Set offset to compensate for probe errors 5 Fig 5 S...

Page 10: ...etooth ON OFF Press the or button to turn on or off see Fig 9 or Fig 10 Fig 9 Fig 10 Note When Bluetooth function is turned ON it is entering standby pairing mode the blue LED built in flashes once ev...

Page 11: ...a specific thermocouple 1 Press the or button to select thermocouple channel see Fig 13 Setting thermocouple channel blinks on the LCD screen 2 Press the button to increase decrease value Note The of...

Page 12: ...or Fig 21 7 3 8 Set system clock The unit is built in a clock so that the data logger function can also record the data and time along with the measurement value 1 Press the or button to select year d...

Page 13: ...to a computer The meter can be connected to PC through a micro USB cable or Bluetooth adapter to download recorded data or perform real time monitoring or datalogging in the software interface 8 POWE...

Page 14: ...9 MAINTENANCE In order to ensure the accuracy of the thermometer for a longer period of time you should calibrate it once a year Clean the device and the window of the display with a clean lint free...

Page 15: ...nnel Thermometer first and connect it to a PC USB port with the cable or set the bluetooth ON 2 Start SE521 program or Click search bluetooth device or to establish connection Note Before clicking sea...

Page 16: ...and file type to save There are three types of file format you can choose binary file ghf text file txt and EXCEL format file csv The ghf file use much fewer disk space to save the data than the other...

Page 17: ...m 6 Choose Data Logger from main menu or click from tool bar 7 In reference to Data Logger see DataLogger 10 7 Main Menu File Open Retrieve files from the disk Save Save the active window when the cap...

Page 18: ...connecting to bluetooth device Disconnect Bluetooth Device Disconnecting from bluetooth device 10 8 DataLogger When you have the thermometer connected to a PC select DataLogger from the main menu or c...

Page 19: ...nse Limited www thermosense co uk Page 18 for examples It will transfer first data set to graph on the right hand side The user can also click at any other data set to choose that set for graph Graph...

Page 20: ...cross sign on the graph This is available only when Split option is unchecked And all the mark you make will exist with T1 When selected the mouse cursor will become a I sign when moving to the graph...

Page 21: ...equently Asked Question 1 How to uninstall SE521 Answer Uninstall SE521 by launching the Add Remove Programs applet out of the Control Panel highlighting the SE521 and clicking on the Add Remove push...

Page 22: ...mited www thermosense co uk Page 21 10 10 Bluetooth Pairing 10 10 1 Via ACTION CENTER Windows 10 1 Open ACTION CENTER 2 Click All Settings 3 Click Devices 4 Select Bluetooth 5 Turn On 6 Select BT xxxx...

Page 23: ...START and then click Devices and Printers 2 Click Add a device 3 Choose BT XXXXX Bluetooth Adapter then click Next 4 To find out which COM port has been allocated to the meter click on the Windows St...

Page 24: ...Thermosense Limited 11 Eghams Court Boston Drive Bourne End Bucks SL8 5YS UK Telephone 44 0 1628 531166 Fax 44 0 1628 531499 Email sales thermosense co uk Web www thermosense co uk...

Reviews: