background image

………………………………………………………………

……………

www.tinytag.info

www.tinytag.info

www.tinytag.info

www.tinytag.info         

sales@tinytag.info

sales@tinytag.info

sales@tinytag.info

sales@tinytag.info    

………………………………………………………………

……………

.

 

Thermocouple

Logger

 

Thermocouple Specific Configuration Options 

Thermocouple Specific Configuration Options 

Thermocouple Specific Configuration Options 

Thermocouple Specific Configuration Options 
in Tinytag Explorer

in Tinytag Explorer

in Tinytag Explorer

in Tinytag Explorer    

    

When starting a thermocouple data logger in Tinytag Explorer you will see two 
options specific to this type of logger that are not mentioned in the software’s 
quick start guide. 
 

Thermocouple Type

Thermocouple Type

Thermocouple Type

Thermocouple Type    
    

Select the thermocouple type you are using from the options listed.   
 
 
 
 
 
 
If you are unsure what type of thermocouple you have, the plug or cable on it may 
be coloured coded and you can match this to the diagram next to each option. 
 
Colours for the different types of thermocouple are as follows: 
 
 
Type K

Type K

Type K

Type K         

    

    

Type J

Type J

Type J

Type J    

    
    
    
Type T

Type T

Type T

Type T         

    

    

Type N

Type N

Type N

Type N    

    

Tinytag Ultra 2 

Tinytag Ultra 2 

Tinytag Ultra 2 

Tinytag Ultra 2     
Thermocouple Manual

Thermocouple Manual

Thermocouple Manual

Thermocouple Manual    

TGU

TGU

TGU

TGU----4550

4550

4550

4550    

Warnings

Warnings

Warnings

Warnings    

 

• 

This logger must be used in accordance with the information provided in this 
manual.   

• 

This equipment contains a lithium battery.  Do not cut open, incinerate, 
recharge or expose to temperatures in excess of 100°C (212°F).  Danger of 
explosion if the battery is incorrectly fitted.  

• 

The battery should only be replaced with the battery type specified in this 
document, observing the correct polarity. 

• 

This equipment should only be connected to a computer using a CAB-0007-
USB interface cable. 

• 

This equipment should be used within the temperature range and other 
environmental conditions specified in this manual. 

Getting Started

Getting Started

Getting Started

Getting Started    

 
To use a Tinytag Ultra 2 thermocouple data logger you will require the following 
items: 
 

• 

A copy of the Tinytag Explorer software 

• 

A CAB-0007-USB download cable

 

 
Install Tinytag Explorer and the USB download cable, as described in the 
software’s quick start guide and plug the USB cable into the socket on the front of 
the logger.   
 
For further information on how to start the data logger recording and how to view 
the results from the unit, please see the Tinytag Explorer quick start guide.

 

Measurement Specification

Measurement Specification

Measurement Specification

Measurement Specification    

 
Thermocouple 

Thermocouple 

Thermocouple 

Thermocouple     

 

Sensor Type 

Sensor Type 

Sensor Type 

Sensor Type  

 

 

Type K, J, T or N Thermocouple 

 

Range

Range

Range

Range   

 

 

Type K

Type K

Type K

Type K         

    

-270 to +1370°C (-454 to 2498°F) 

 

Type J

Type J

Type J

Type J         

    

-210 to +1200°C (-346 to 2192°F) 

 

Type T

Type T

Type T

Type T         

    

-270 to +400°C (-454 to 752°F) 

 

Type N

Type N

Type N

Type N         

    

-270 to +1300°C (-454 to 2372°F) 

Reading Resolution 

Reading Resolution 

Reading Resolution 

Reading Resolution  

 

0.01°C  

Cold Junction Compensation

Cold Junction Compensation

Cold Junction Compensation

Cold Junction Compensation 

-10 to +70°C

 

Accuracy

Accuracy

Accuracy

Accuracy        

    

    

Better than ±1.0°C across all thermocouple  

 

 

 

 

ranges when the logger is between   

 

 

 

 

-10 & +70°C. 

 
Note: The above accuracy figures do not include the thermocouple probe.    

 

Logger Temperature

Logger Temperature

Logger Temperature

Logger Temperature    

 
Range

Range

Range

Range   

 

 

-40 to +85°C  (-40 to +185°F) 

Sensor Type 

Sensor Type 

Sensor Type 

Sensor Type  

 

 

10K NTC Thermistor  

 

 

Response Time

Response Time

Response Time

Response Time   

 

20 mins to 90% FSD in moving air 

Reading Resolution 

Reading Resolution 

Reading Resolution 

Reading Resolution  

 

0.02°C or better 

Accuracy

Accuracy

Accuracy

Accuracy        

    

    

Better than ±0.5°C 

Burnout Response

Burnout Response

Burnout Response

Burnout Response    

 
The burnout response option allows you to tell the data logger what to do in the 
event of a problem with the thermocouple being used. 

If a faulty thermocouple is attached to the data logger it is possible for the logger 
to show readings that may appear to be correct.  The burnout response can 
deliberately drive readings low or high in the event of a problem, to make it clear 
to the user that there is a problem. 
 
Choose what you want the data logger to do from the options listed. 

Using the Thermocouple Data Logger

Using the Thermocouple Data Logger

Using the Thermocouple Data Logger

Using the Thermocouple Data Logger    

 
To get the best results, ensure that the data logger and the thermocouple plug 
have sufficient time to reach the same temperature before recording begins. If the 
thermocouple has been stored at a significantly different temperature from the 
logger, allow 30 minutes for stabilisation.   
 
It is advisable to keep the logger at a constant temperature whilst recording, and 
to avoid large temperature gradients across the unit (for example, avoid placing 
the logger on a very hot or very cold surface).   
 
Large fluctuations in logger temperature can cause temporary errors in the probe 
measurement, although these are unlikely to exceed 1°C. 

Reviews: