background image

Wiper W55 V2

1

TriOS Mess- und Datentechnik GmbH · Bürgermeister-Brötje-Str. 25 · D-26180 Rastede · Germany 
fon: +49 (0) 4402 69670 - 0 · fax: +49 (0) 4402 69670 - 20 · info@trios.de · www.trios.de

D04-067en202111 Quick Guide Wiper W55 V2

Wiper W55 V2

02A100XX8

Technical Specifications

Path lengths                       

2 mm, 5 mm, 10 mm

Control port

4-pin M8-plug 

A suitable M8 connection cable with open end is included in the scope of 

delivery.

Trigger input

5-24 VDC (±10% )

Power consumption  
trigger input

2...15 mA

Operating time (max.)                     

3 Seconds

Dimensions L x Ø

175 mm x 80 mm

Weight

0.52 kg

Material

NBR, POM, TPE (PP, EPDM), Titanium or V4A

Power supply

12-24 VDC (± 10 %)

Power consumption

approx. 2-6 W in operation; max. 0.75 W in standby

Maintenance effort

≤ 0.5 h/month typical

Maintenance interval

depending on application

Warranty

1 year (EU & USA : 2 years)

Max. Pressure

1 bar

Protection Type

IP68

Inflow velocity

up to 10 m/s

Operating temperature

+2...+40 °C

Storage temperature

-10 °C...+70 °C                                    

The TriOS Wiper W55 V2 provides an additional cleaning option for all TriOS photometers with path lengths 
up to 10 mm. The wiper housing can be mounted on the sensor in just a few steps and provides reliable 
cleaning of the measurement windows. The new magnetic axis lock allows quick and easy wiper blade re

-

placement, without any tools.

The new version of the wiper now features blockage detection and removal, and a service mode that increas

-

es the life of the wiper through regular use. The accessory can also be used in seawater up to a depth of 10m. 

Summary of Contents for 02A100 8 Series

Page 1: ...2 6 W in operation max 0 75 W in standby Maintenance effort 0 5 h month typical Maintenance interval depending on application Warranty 1 year EU USA 2 years Max Pressure 1 bar Protection Type IP68 In...

Page 2: ...0 The wiper blades and the axis are not included in the scope of delivery of the wiper housing as these must be selected depending on the path length of the sensor A set of wiper blades contains five...

Page 3: ...spring in the wiper housing The axis groove is magnetically tightened and must be inserted as far as it will go and if necessary turned slightly until the correct position engages Wiper blades and ax...

Page 4: ...the wiper housing The axis groove is magnetically tightened and must be inserted as far as it will go and if necessary turned slightly until the correct position engages 3 Change housing seal O ring...

Page 5: ...r an alternative connection where power supply and cleaning control triggering are separate the open end of the M8 line can be disconnected see option 2 Remove the twin ferrules and equip the individu...

Page 6: ...nd unscrewing the four screws under neath 2 Place the wiper cable through the cable guide of the TriBox mini and tighten the cable guide 3 Connect the blue black wire of the wiper cable to the CON2 co...

Page 7: ...ng the screws underneath TriBox3 Wiper Open End Cable 2 Place the wiper cable through the TriBox3 cable guide and tighten the cable guide on the wider part of the cable 3 Connect the blue black wire o...

Page 8: ...ed magnet and five full rotations in both directions are performed This adjusts the home position and lubricates the planetary gear in all directions It is recommended to remove the upper wiper housin...

Page 9: ...per protective cage 00P100010 Size L x 220 mm x 88 mm Weight 0 5 kg Material POM Technical Specifications The sturdy plastic POM protective cage has been designed to keep coarse dirt and larger object...

Page 10: ...of the protective cage on the sensor 1 The wiper must be correctly installed before mounting the cage on the sensor see page 2 to 4 2 Push the protective cage with the opening in the direction of the...

Page 11: ...urge on the wiper 1 The adjustment screw needs to be fixed to the lower housing for mounting the compressed air purge Spare part no 10P000000 2 Moreover the compressed air fitting needs to be inserted...

Reviews: