background image

PAGE 1 OF 8 

0422 IH-9884

1-800-295-5510

uline.com

H-9884, H-9885

H-9886

BALANCE SCALE

OVERVIEW OF CONTROLS

CONTROL PANEL

NET

No

Units

g

Exit

Back

Mode

Menu-Cal

Yes

Zero

Print

Function

Tare

CONTROL FUNCTIONS

BUTTON

FUNCTION

Short Press (when on): Sets display to zero. 

                  (when off): Turns scale on. 

 

Long Press (when on): Turns the scale off. 

Short Press (in Menu): Selects/accepts displayed setting.

Long Press: Toggles through active units. 

 
Short Press (in Menu): Toggles through available settings.

Short Press: Selects function setting. 

Long Press: Selects active mode. 

 

Short Press (in Menu): Returns to previous settings.

Short Press: Enters/clears a Tare value. 

Long Press: Enters user menu.

Short Press (in Menu): Quickly exits user menu.

Zero

Yes

Print

Units

No

Function

Mode

Back

Tare

Menu-Cal

Exit

Summary of Contents for H-9884

Page 1: ...e on Long Press when on Turns the scale off Short Press in Menu Selects accepts displayed setting Long Press Toggles through active units Short Press in Menu Toggles through available settings Short P...

Page 2: ...Disconnect from power supply when cleaning Service should only be performed by authorized personnel SETUP TRANSPORTATION LOCK 1 Rotate the pointer on the transportation lock to the unlocked position S...

Page 3: ...4 After pressing yes the scale will then display the stored sample size Press NO or BACK to toggle the choices 5 10 20 50 or 100 5 Put the indicated number of pieces on the pan Press YES to calculate...

Page 4: ...on procedure zero and span A span calibration is important when initially setting up the scale Lin LIN yes no Initiates a linearity calibration procedure zero mid point and span SETUP SUB MENU Back Li...

Page 5: ...he menu must be turned on to be active NOTE Available units and modes vary by model and local regulations SECURITY SWITCH Menu Lock switch may be used to limit changes to the user menu See Figure 4 Th...

Page 6: ...nge exceeded Clear pan check shipping lock setting Err 8 2 LoAd Power on zero under range Install pan check shipping lock setting Err 8 3 LoAd Overload Load exceeds scale maximum capacity Err 8 4 LoAd...

Page 7: ...des Weighing Parts Counting Checkweighing Power Requirement AC adapter included or four C batteries not included Typical Battery Life 270 hours 200 hours 270 hours Specified Temperature Range 50 F to...

Page 8: ...0 0001 21 574 x 0 001 Ounce oz 7 7600 x 0 0005 77 6025 x 0 0005 77 600 x 0 005 Ounce Troy ozt 7 0730 x 0 0005 70 7315 x 0 0005 70 730 x 0 005 Pennyweight dwt 141 46 x 0 01 1 414 63 x 0 01 1 414 6 x 0...

Reviews: