background image

Version 1 

 

 

 

 

 

 

 

  ULTRAPROBE

®

 201

 

Instruction Manual 

 

 

Summary of Contents for ULTRAPROBE 201

Page 1: ...Version 1 1 ULTRAPROBE 201 Instruction Manual ...

Page 2: ...ouching the equipment with your hands or fingers Besure to use appropriate lockout procedures when attempting repairs Be careful with loose hanging parts such as the wrist strap or headphone cord when inspecting near moving mechanical devices since they may get caught Don t touch moving parts with the contact probe This may not only damage the part but cause personal injury as well When inspecting...

Page 3: ...c components 6 Component Assembly Instructions 7 UE Systems Grease Caddy Docking Station Mounting Instructions 7 UE Systems Grease Caddy Docking Station Mounting Instructions 8 Metered Housing 9 Headphones 10 Using the Ultraprobe 201 Grease Caddy 11 Specifications Ultraprobe 201 Grease Caddy 12 ...

Page 4: ...e the amplitude of the sound on ameter The short wave high frequency nature of ultrasound offers many advantages for instruments used to inspect operating equipment 1 The amplitude of generated ultrasounds will fall off rapidly as the sound travels from its source providing a localized signal that can be detected and monitored with relative ease in most noisy environments 2 The subtle nature of th...

Page 5: ... increase in low frequency vibration levels Ultrasonic inspection of bearings is useful in recognizing the beginning stages of fatigue failure related to flooding of or lack of lubricant In ball bearings as the metal in the raceway roller or ball bearing begins to fatigue a subtle deformation begins to occur This deforming of the metal will produce irregular surfaces which will cause an increase i...

Page 6: ...sion 1 6 Component assembly magnetic mount sensor Basic components Your kit consists of the following parts A Grease Caddy Assembly B Clamp C Sensor Assembly Magnetic Sensor and Cable with Acoustic Shield ...

Page 7: ... to familiarize yourself with the basic components of your kit UE Systems Grease Caddy Docking Station Mounting Instructions Docking Station Note the Raised disc with the logo This is to face away from the grease gun 1 Remove tip of grease gun 2 Line up indent facing grease gun 3 Srew on Docking Station and tighten 4 Screw on grease gun tip and tighten NOTE Do not use Docking Station when using ru...

Page 8: ... Systems Grease Caddy Docking Station Mounting Instructions Completed setup Close up view Ready to use Keep Magnetic on Docking Station when greasing NOTE Do not use Docking Station when using rubber or plastic tubes ...

Page 9: ...eight 8 sensitivity levels which read out in related decibels of 0 to 70 As the dial is turned to the right to 0 the sensitivity of the instrument increases As the dial is turned to the left to 70 the sensitivity decreases A low level ultrasound emission produces low amplitude To detect low level ultrasounds the instrument should be in a highsensitivity position 0 is the highest sensitivity positi...

Page 10: ...or use in high noise areas These heavy duty headphones are designed to block out intense sounds often found in industrial environments so that the user may easily hear the sounds received by the ULTRAPROBE 201 b For those situations in which it is not possible or difficult to wear the standard headphones described above UE Systems has two other options available 1 the DHC 1991 Earpiece which loops...

Page 11: ...n a Be sure the greasecoupler is securely connected to thezerk fitting and that the magnetic sensor is in full contact with the bearing housing b Wear your headphones and besure the head phones are plugged into the headphone jack c Turn the Ultraprobe 201 ON d If the sound is too loud reduce the sensitivity turn the Sensitivity Dial LEFT until the LED s are at approximately 50 of scale e Apply the...

Page 12: ...requency Response Peak response centered around 38kHz Indicator 10 Segment LED Bar Graph red Bright white LED for illumination of test area Sensitivity Selection 8 position precision attenuation Power Rechargeable nickel metal hydride Power off Time delay after five 5 minutes Low Batteryindicator LED Headset Deluxe noise isolating headset for hard hat use Over 23 dB of noise attenuation Meets or e...

Page 13: ...ed further support Want information regarding products or training Contact UE Systems Europe Windmolen 22 7609 NN Almelo NL e info uesystems eu w www uesystems eu t 31 0 548 659 011 f 31 0 548 659 010 www uesystems eu ...