background image

neon

3™

universal OCT spinal stabilization

Assembly and disassembly instructions 

with special cleaning instructions

Summary of Contents for neon3 CS 3921 Series

Page 1: ...neon3 universal OCT spinal stabilization Assembly and disassembly instructions with special cleaning instructions ...

Page 2: ... in the component overview are listed in English according to the labeling on the instruments The user must ensure that the current versions of the system related surgical technique and the Processing manual implants and instruments UH 1100 are on hand and are followed These are also available at www ifu ulrichmedical com Labeling and symbols Brush Rinse Move back and forth several times Click sou...

Page 3: ...for CS 3929 XX neon3 Assembly and disassembly instructions for CS 3930 XX neon3 Assembly and disassembly instructions for CS 3932 XX neon3 Assembly and disassembly instructions for CS 3938 03 04 05 neon3 Assembly and disassembly instructions for CS 3938 07 08 09 neon3 Assembly and disassembly instructions for CS 3954 01 neon3 Assembly and disassembly instructions for CS 3955 02 Contents ...

Page 4: ...neon3 Assembly and disassembly instructions for CS 3921 XX Components involved No Product no Product description 1 CS 7150 Gearshift handle coupling B 2 CS 3921 01 Awl cannulated for screws Ø 4 5 5 5 mm coupling B 3 CS 3921 02 Trocar wire for awl thoracic ...

Page 5: ... screwed in While inserting the awl the release on the gearshift handle must be pulled backwards Disassembly instructions All the individual parts of the awl with trocar wire must be disassem bled for cleaning and storage on the tray Disassembly of the cannulated awl The trocar wire is unscrewed from the awl and pulled out to the rear By pulling the release the awl can be removed from the gearshif...

Page 6: ...ere is any visible soiling If this is the case the previous steps must be repeated Maintaining the awl with trocar wire For their maintenance movable parts such as joints hinges and threaded parts should be treated with suitable maintenance oil In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply neon3 Assembly and disassembly instructi...

Page 7: ...h stop adjustable 3 CS 3928 04 Trocar cannulated for drill 2 4 CS 3927 04 Adapter for drill 2 serrated 5 CS 3926 04 Adapter for drill 1 blunt 6 CS 3926 05 Adapter for drill 1 serrated 7 CS 3923 01 Trocar cannulated for trocar tipped drill wire Ø 1 5 mm 8 CS 3929 06 Reference star for use with Brainlab neon3 Assembly and disassembly instructions for CS 3929 XX ...

Page 8: ... 04 CS 3926 05 CS 3927 04 are attached to the drill guide CS 3929 05 until it engages Attaching the reference star The push button on the reference star for use with Brainlab CS 3929 06 must be held down so that the reference star can be placed onto the drill guide CS 3929 05 Inserting the depth stop The release of the drill guide CS 3929 05 is turned clockwise Then the depth stop CS 3929 02 can b...

Page 9: ...ll guide They can be removed from the drill guide by pulling on the drill attachment Removing the depth stop The release of the drill guide is turned clockwise Then the depth stop can be pulled out of the drill guide Removing the reference star The push button on the reference star for use with Brainlab CS 3929 06 must be held down so that the reference star can be pulled off of the drill guide ...

Page 10: ... of professionally Dismantling the drill guide Difficult to reach parts must be pre cleaned manually with a brush in a water bath In addition the release should be brushed when both opened and closed and moved back and forth several times The instrument must then be rinsed thoroughly to avoid any residue from drying neon3 Assembly and disassembly instructions for CS 3929 XX ...

Page 11: ...ment must then be rinsed thoroughly to avoid any residue from drying Visual inspection Finally it must be checked whether there is any visible soiling If this is the case the previous steps must be repeated Components maintenance For the maintenance of the com ponents movable parts threads hinges and buttons should be treat ed with suitable maintenance oil In addition the general cleaning in struc...

Page 12: ... cleaning Difficult to reach parts must be pre cleaned manually with a brush in a water bath The instrument must then be rinsed thoroughly to avoid any residue from drying Visual inspection Finally it must be checked whether there is any visible soiling If this is the case the previous steps must be repeated In addition the general cleaning instructions of the Processing manual implants and instru...

Page 13: ...d No Product no Product description 1 CS 3932 04 Screwdriver cannulated with SR 15 2 CS 3932 02 Sleeve for screwdriver 3 CS 3932 05 Rotating end for CS 3932 04 4 CS 3932 07 Screwdriver cannulated with SR 15 coupling B neon3 Assembly and disassembly instructions for CS 3932 XX ...

Page 14: ...ctions for CS 3932 XX Disassembly instructions The screwdriver must be disassembled for cleaning and storage on the tray Removing the sleeve By pulling on the sleeve it can be removed from the screwdriver Removing the rotating end The rotating end is pulled off of the handle to the rear ...

Page 15: ...ust then be rinsed thoroughly to avoid any residue from drying Important All lumens and cavities must be rinsed out carefully and thoroughly Visual inspection Finally it must be checked whether there is any visible soiling If this is the case the previous steps must be repeated Maintaining the screwdriver For their maintenance movable parts such as joints hinges and threaded parts should be treate...

Page 16: ...bly and disassembly instructions for CS 3938 03 04 and 05 Components involved No Product no Product description 1 CS 3938 03 Star grip handle for rod pusher Persuader 2 CS 3938 04 Inner sleeve for rod pusher Persuader 3 CS 3938 05 Outer sleeve for rod pusher Persuader ...

Page 17: ...hed the inner sleeve turns clockwise and after releasing it it snaps securely into the outer sleeve For better handling a star grip handle CS 3938 03 can optionally be attached to the rod pusher Disassembly instructions The rod pusher must be disassembled for cleaning and storage on the tray Removing the outer sleeve If the inner sleeve is turned to the left when in a released state and with the a...

Page 18: ... the end of the outer sleeve should be brushed The instrument must then be rinsed thoroughly to avoid any residue from drying Cleaning the inner sleeve Difficult to reach parts should be pre cleaned manually with a brush in a water bath The spring and adjusting nut must be moved back and forth several times In addition the cannulation and the adjusting nut should be rinsed while being constantly t...

Page 19: ...iling If this is the case the previous steps must be repeated Components maintenance For the maintenance of the components movable parts threads hinges and buttons should be treated with suitable maintenance oil In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply ...

Page 20: ...ly and disassembly instructions for CS 3938 07 08 and 09 Components involved No Product no Product description 1 CS 3938 07 Handle for rod reduction instrument 2 CS 3938 08 Outer sleeve for rod reduction instrument 3 CS 3938 09 Inner sleeve for rod reduction instrument ...

Page 21: ...until the circumferential groove of the inner sleeve is flush with the end of the outer sleeve Moreover the outer sleeve and the inner sleeve are aligned in such a way that the respective longitudinal grooves line up see diagram The handle of the rod reduction instrument CS 3938 07 is completely compressed and the previously assembled sleeves are inserted onto the prongs on the handle for the rod ...

Page 22: ...torage on the tray Removing the sleeves The handle on the rod reduction instrument is completely compressed The inner sleeve and the outer sleeve are aligned in such a way that the respective longitudinal grooves line up see diagram The sleeves can now be pulled off of the handle from above After the outer sleeve and the inner sleeve have been removed from the handle of the rod reduction instrumen...

Page 23: ...ing In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply Cleaning the outer sleeve Difficult to reach parts should be pre cleaned manually with a brush in a water bath In addition the notch at the end of the outer sleeve of the rod reduction instrument should be brushed The instrument must then be rinsed thoroughly to avoid any residue ...

Page 24: ...ng manual implants and instruments UH 1100 also apply Visual inspection Finally it must be checked whether there is any visible soiling If this is the case the previous steps must be repeated In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply Components maintenance For the maintenance of the components movable parts threads hinges and...

Page 25: ... 3954 01 OC plate bending clamp Special cleaning instructions Cleaning the OC plate bending clamp Difficult to reach parts should be pre cleaned manually with a brush in a water bath In addition the release should be brushed when both opened and closed and moved back and forth several times The instrument must then be rinsed thoroughly to avoid any residue from drying ...

Page 26: ...hether there is any visible soiling If this is the case the previous steps must be repeated Components maintenance For the maintenance of the components movable parts threads hinges and buttons should be treated with suitable maintenance oil In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply ...

Page 27: ...ng coupling AO neon3 Assembly and disassembly instructions for CS 3955 02 Special cleaning instructions Manual cleaning Difficult to reach parts must be pre cleaned manually with a brush in a water bath When doing so the working end should be moved back and forth several times The instrument must then be rinsed thoroughly to avoid any residue from drying ...

Page 28: ...e checked whether there is any visible soiling If this is the case the previous steps must be repeated Components maintenance For the maintenance of the components movable parts threads hinges and buttons should be treated with suitable maintenance oil In addition the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply neon3 Assembly and disassembly i...