background image

Instructions for use

301 plus

Summary of Contents for assistina 301 plus

Page 1: ...Instructions for use 301 plus...

Page 2: ...ng of the system 14 8 Test 15 9 Function check Function display function check of the mechanical drive 16 10 Maintenance 17 Maintenance of transmission instruments with ISO coupling 18 Maintenance of...

Page 3: ...t identification e g in hygienic maintenance process WARNING Risk of injury ATTENTION To prevent damage occurring General explanations with no danger to persons or property Highly flammable Avoid sour...

Page 4: ...ning the spray channels and lubricating the moving parts of dental transmission instruments turbines air motors and air driven dental scalers Misuse can damage the Assistina and involves risk and dang...

Page 5: ...ternalattachmentsorthelike Responsibility of the manufacturer W H Dentalwerk B rmoos accepts responsibility for consequences affecting the safety reliability and performance of the Assistina only when...

Page 6: ...e original packaging Have the User Manual at hand Take out the white and green funnels Remove the turbine adaptor LiftouttheAssistinainthefibre board box Open the box at the tucked in closureandremove...

Page 7: ...filter complete 4 hole turbine adaptor REF 02685000 Funnel white REF 02689300 Funnel green REF 02689400 Air connection hose 2 m REF 02697000 complete Initial provision of 5 layers of fleece REF 02690...

Page 8: ...re starting up the unit Keep cleaning fluid service oil away from the eyes Do not swallow cleaning fluid Keep service oil away from the skin Cleaning solution and its residues are flammable Flash poin...

Page 9: ...uid Indicator window for service oil Fill level of service oil Waste disposal tray Recess for adaptor Cover Filler hole storage container for cleaning fluid Filler hole storage container for service o...

Page 10: ...10 6 Description of the rear side Suction filter Quick release coupling for compressed air...

Page 11: ...them in a dust free environment Check the filling level indicator Use only clean filtered compressed air The W H cleaning fluid is aldehyde free and nontoxic It is biodegradable in waste water W H ser...

Page 12: ...Unscrew the cap from the white storage container Insert the white funnel Add 250 ml of service oil Unscrew the cap from the green storage container Insert the green funnel Add 250 ml of cleaning fluid...

Page 13: ...It cannot clean unfiltered and possibly contaminated compressor air Dirty fine filters must be replaced Fit the suction filter so that it is leak tight Push the quick release coupling of the connecti...

Page 14: ...ol the supply lines are not yet completely full Pull the drip guard off the maintenance attachment and store it safely Hold a cloth or tissue in front of the maintenance attachment Press the program b...

Page 15: ...tachment remove it Check that the fleece inlay is damp to ensure correct function If it is not damp repeat the process 2 to 3 times Dispose of saturated fleece inlays as with cloths for surface disinf...

Page 16: ...balls will drop down again Correct operation of the maintenance process is indicated by these two function displays If ever the balls disappear during a program cycle this indicates a malfunction Fun...

Page 17: ...the manufacturer has allowed for removal of the treatment heads With sterilization after every patient Press the program button once With maintenance twice a day Press the program button twice on eac...

Page 18: ...se button and pull off the turbine adaptor Keep the adaptor in the recess With its spray exit nozzle directed downwards push the transmission instrument onto the maintenance attachment up to the limit...

Page 19: ...for use with turbine hose Other turbines Screw the turbine adaptor to the relevant quick release coupling e g adapter 02692000 04713200 Turbines with 2 hole or 3 hole Borden connection and spray air t...

Page 20: ...onto the adaptor until it engages Close the cover and the waste disposal tray Push the appropriate adaptor onto the maintenance attachment up to the limit stop Press the program button once for about...

Page 21: ...will be completed in about 35 seconds Use the turbine adaptor REF 02685000 with standard 4 hole connection Set the rotary ring of the air motor to clockwise or anticlockwise running no maintenance wi...

Page 22: ...ected downwards push the T1 adaptor onto the maintenance attachment Initially keep the cover open Push the instrument onto the adaptor from underneath Support the instrument at the edge of the cover P...

Page 23: ...s required Use W H cleaning fluid to clean the outside of the case Clean the cover and the waste disposal tray as required Squeeze the cover slightly Remove the cover from its fixing Use W H cleaning...

Page 24: ...690400 Adaptor for all W H products with Roto Quick system 02691000 Adaptor for Sirona T1 Classic transmission instruments 02692000 Adaptor for Sirona T1 T2 T3 turbines 02693000 Adaptor for cleaning F...

Page 25: ...000 02695700 O ring for maintenance attachment 02670500 Filter pack 02675910 500 ml W H service oil 02680200 1000 ml W H cleaning fluid 02680500 Pack of W H service oil and W H cleaning fluid 02690300...

Page 26: ...ly positioned will ensure consistently safe and correct maintenance Once a year Carry out a function check Page 16 Check the fine filter in the air connection hose for contamination Check the O rings...

Page 27: ...ttached here The O rings can be exchanged without the need for special tools Grease free fingers will facilitate this exchange Firmly squeeze the O ring between thumb and forefinger to form a loop Pul...

Page 28: ...he suction filter after approx 7 000 maintenance cycles corresponds to the consumption of a 500 ml bottle of W H service oil Pull out the suction filter horizontally towards the rear Press out the fil...

Page 29: ...ou have any questions please contact your local dental dealers or one of our service points see Page 33 Always use the original packaging for return transport Service 2 transport seals No 02680600 are...

Page 30: ...ainers simultaneously The cleaning fluid and service oil must not be mixed Fit a seal in each filler hole Screw on the two caps Before emptying the first container fit the transport seal on the second...

Page 31: ...ption approx 60 l min Noise 63 dB A at 50 cm distance Flow pressure 4 bar to 10 bar pressure regulation provided by integral automatic pressure regulator Capacity 250 ml service oil sufficient for app...

Page 32: ...ble for material or manufacturing defects within a warranty period of 12 months from the date of purchase Weacceptnoresponsibilityfordamagecausedbyincorrecthandlingorbyrepairscarriedoutbythirdpartiesn...

Page 33: ...Centre Park Street St Albans Herts AL2 2NJ United Kingdom t 44 1727 874990 f 44 1727 872254 E Mail technical uk wh com W H Impex Inc 6490 Hawthorne Drive Windsor Ontario N8T 1J9 Canada t 1 800 2656277...

Page 34: ......

Page 35: ......

Page 36: ...Manufacturer Form Nr 50458 AEN Rev 016 15 06 2016 Subject to alterations W H Dentalwerk B rmoos GmbH Ignaz Glaser Stra e 53 5111 B rmoos Austria t 43 6274 6236 0 f 43 6274 6236 55 office wh com wh com...

Reviews: