background image

 

6-16-19, JINGUMAE, SIBUYA-KU, TOKYO 150-0001, JAPAN 

TEL: +81-3-3400-6147    FAX: +81-3-3409-3156

 

 

1/2 

IM0789-01 

INSTRUCTION 

MANUAL

 

WSP-RTS 

RTD SIGNAL CONVERTER 

 

 

Thank  you  for  selecting  another  fine 

  product.  Please 

check  the  description  given  on  the  rating  label  of  this  unit  to  make  sure 
that it meets your specifications and be sure to read this instruction manual 
before using the product. 
This manual outlines the operation, connection and adjustment procedures 
of this product. 
The  unit  has  been  manufactured  and  inspected  according  to  our  strict 
quality  control  standard.  If  you  should  find  a  defect  including  damage 
incurred  during  transportation,  contact  us  or  the  dealer  where  you 
purchased it immediately. 

○PACKAGE INCLUDES:   

RTD signal converter.…………………1 

Base socket……………………………1 

 
* For details of models and specifications, please download the product 

catalog from our website, and then check it. 

 

1. PRECAUTIONS 

1) Conformity with CE directive 

 

This  equipment  is  compliant  with  Installation  Category  II  and 

Pollution  Degree  2  environment.  The  insulation  capability  between 

signal  input  and  output  is  basic  insulation.  Before  installing,  please 

check that the insulation class of this equipment satisfies your system 

requirements. 

 

Please be sure to install this equipment to the inside of a panel. 

 

Please use this equipment at an altitude of up to 2000 m. 

 

Compliance with EN standards:   

   

EN61326-1 

   

EMS: Industrial environments; EMI: Class A 

The wiring length should be not more than 30 m. 

EN61010-1 

EN IEC 63000 

*  Custom-made  items other  than  a  catalog  standard  specification  are 

outside CE conformity. 

2) Power supply 

 

Check  the  rated  voltage  on  the  rating  label,  and  use  the  product 

within the range of each of the following ratings (

  is indicated on 

the rating label.) 

 

100 to 240VAC 

100 to 240VAC ±10% (50/60 Hz) approx. 4.5 VA 

 

24VDC 

24VDC ±10% approx. 66 mA 

* You can use it as 10.8 to 30 V, if you do not use this product as a   

CE compliant article.   

 

110VDC 

110VDC ±10% approx. 13mA 

3) Handling 

 

When  removing  or  mounting  the  main  body  from/to  the  socket,  be 

sure to turn OFF the power supply and the input signal in advance to 

prevent any problems. 

4) Installation 

 

This equipment is designed for indoor use. 

 

Please  install  the  main  body  in  a  location  where  the  ambient 

temperature is within -5 to 55°C. 

 

Please install the main body in a location where relative humidity is 

less than 90%RH (no freezing or condensation). 

 

When  the  equipment  is  to  be  installed  in  a  location  where  there  is 

excessive  dust  or  metal  particles,  house  it  in  a  dust-proof  cabinet, 

which has a heat radiation function. 

 

Avoid  exposing  the  equipment  to  vibration  and  impact,  which  may 

cause malfunction. 

 

Please do not block ventilation openings of the main body. 

5) Wiring 

 

Be  sure  to  keep  the  wiring  of  the  power  line,  input  signal  line  and 

output signal line away from any noise source, relay driving line and 

high-frequency line. 

 

Avoid  clamping  these  lines  together  with  a  noise-superimposed  line 

or putting them together in the same duct. 

6) Others 

 

This  equipment  can  be  operated  as  soon  as  the  power  supply  is 
turned ON. However, for optimum performance, allow 30 minutes of 
energizing time. 

 

This  product  is  precision  equipment.  It  is  therefore  recommended 
that  you  periodically  calibrate  it  by  performing  ZERO  and  SPAN 
adjustments once a year. 

 

2. OUTLINE DIMENSIONS 

 
 
 
 
 
 
 
 
 
 
 
 

3. CIRCUIT DIAGRAM 

 
 
 
 
 
 
 
 
 
 

 

 

4. NAME OF EACH PART AND 

TERMINAL LAYOUT 

 
 
 
 
 
 
 
 
 
 

 

Terminal No.

 

Symbol

 

Contents

 

INPUT 

Please connect a prescriptive input signal. 

NC

 

No Connection (

1) 

OUTPUT-1 

 

The signal based on the input/output 
specification is output.

 

12 

 

13 

POWER 

U(+) 

Connect  the  power  supply  of  the  rated 
voltage.

 

14 

 

V(-) 

Connection points 

M3 SEMS screw 

Screw tightening torque 

Recommended: 0.6 N-m 

1 Please do not wire anything at NC terminal. 

Unit:mm

6.5

84

22

100

70

30

DIN rail

35mm

width

4

SPAN

ZERO

OUT1

Rating

label

Terminal

label

MODEL

TAG NO.

WSP-

2-

φ

4.2×5

Mtg hole

72

15

23

59

Base socket

6

8-M3

Screw

Rating label

Terminal label

Main unit

Socket(8P)

OUTPUT-1

 Span-trimmer
 Zero-trimmer

Setscrews of

the main unit

12

14

9

13

1

8

5

4

 U(+)
 V(-)

POWER

OUTPUT-1

 -

 +

 B
 A  INPUT

 B

Model label

Constant current circuit

Burnout circuit

Burnout circuit

Voltage amplifier

circuit

Digital conversion

circuit

Isolation circuit

ZERO

The first

output
circuit

SPAN

2

Power circuit

 (AC/DC)

1

Standard resistance

1

4

5

8

9

12

13

14

B

RTD

B

A

OUTPUT-1

(Base socket assembly)

U(+)

V(-)

POWER

NC

Reviews: