background image

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Installation, Operation and  
Maintenance Manual 

Lync WQ-UV

 

Complete Water Quality Solutions

 

Engineered Solutions 

 Lync

  

•  425 W Everman Pkway, St. 101  •  Fort Worth, TX 76134 

USA:

 T: (817) 335-9531  

Technical Support

  

•  (800) 433-5654 (ext. 3)  •  Mon-Fri, 8 am - 5 pm CST 

© 2021 LYNC 

L- OMM-006_0A

    

    

1/26/2022

 

 

Disclaimer 

The information contained in this manual is subject to change without notice 
from Lync, a division of Watts Water Technologies, Inc. Lync makes no 
warranty of any kind with respect to this material, including, but not limited to, 
implied warranties of merchantability and fitness for a particular application. 
Lync is not liable for errors appearing in this manual, nor for incidental or 
consequential damages occurring in connection with the furnishing, 
performance, or use of these materials. 

 

Summary of Contents for Lync WQ-UV

Page 1: ...The information contained in this manual is subject to change without notice from Lync a division of Watts Water Technologies Inc Lync makes no warranty of any kind with respect to this material incl...

Page 2: ...nication 19 3 7 1 Serial Communication Settings 19 3 7 2 Supported Function Codes 19 3 7 3 Input Registers 19 3 7 3 1 Lamp Type 21 3 7 3 2 Lamp Rated Operating Time 22 3 7 3 3 Controller Operating Tim...

Page 3: ...UV light penetrates the cell walls of microorganisms and disrupts their genetic deoxyribonucleic acid DNA material This quickly inactivates microorganisms by destroying their ability to replicate and...

Page 4: ...of experience and knowledge if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved Children shall not play with the appli...

Page 5: ...rains 120 mg L as CaCO3 Max iron 0 3 ppm Max manganese 0 05 ppm Turbidity Total suspended solids 5 NTU 10 mg L Oil and H2S Not allowed Dimension W x D x H inches 48 x 24 x 70 60 x 24 x 70 Shipping wei...

Page 6: ...screen auto reset with new lamp Visual alarm Animated icons and text messages on graphic display Audible alarm audible alarm mute Yes yes Alarm relay Solenoid contacts Yes 24V 5 Amp Maximum Alarm rel...

Page 7: ...2 WQ UV Components L OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 7 of 38 2 WQ UV Components 2 1 Skid Components WQUV 050 D WQUV 100 D Figure 2 1 General WQUV 100 D Compon...

Page 8: ...ing and wiring to minimize installation time and labor Components are fastened onto a steel skid for safe and easy transportation and installation 1 Confirm that the system being installed matches the...

Page 9: ...n 10 feet of the system 6 Connect the outlet of the skid to treated water line plumbing If no isolation valve exists it is recommended to install an isolation valve within 10 feet of the system WARNIN...

Page 10: ...be supplied through a residual current device RCD having a rated residual operating current not exceeding 30 milliamps WARNING This appliance must be grounded In the event of a malfunction or breakdow...

Page 11: ...re Automatic lamp dimming for cooler operation Flow switch Lamp out audible alarm Glow cap lamp indicator Graphic touch screen display to show key operational data Alarm output for solenoid valve UV s...

Page 12: ...rm Up Screen the Home Screen will automatically appear This is the screen that the system will display during normal operation From the Home Screen four sub screens can be accessed by touching the ico...

Page 13: ...ort 800 433 5654 Mon Fri 8 am 5 pm EST 13 of 38 4 Set Up Screen Upon touching the screen allows the user to select a language as well as the bypassing and silencing of alarms Touching the house icon f...

Page 14: ...s Upon start up of the system if the UV sensor is not found the user will be notified via this screen The system will continue to the Home Screen when a UV sensor is not found and display the warning...

Page 15: ...the lamp to be in a dimmed condition to reduce water temperature and conserve energy If the system detects low UV intensity a yellow warning will appear If the UV intensity is too low for safe operati...

Page 16: ...o the controller Clean or replace sensor if needed 3 6 5 Set Up Screen The Set Up Screen can be accessed by pressing the gear icon on the Home Screen This allows the user to select a language bypass t...

Page 17: ...te the audible alarm for 24 hours and return to the Set Up Screen Touch No to return to the Set Up Screen with the alarm not muted Audible Alarm is muted Audible Alarm is not muted 3 6 8 Lamp and Slee...

Page 18: ...5654 Mon Fri 8 am 5 pm EST 18 of 38 Touching the QR Code icons on the Lamp and Sleeve Replacement Screen will enlarge them If there are no buttons touched over a 30 seconds period the system will go...

Page 19: ...6 Report Server ID 0x11 17 3 7 3 Input Registers The MODBUS input registers of the Lync UV Controller are used to query information about the controller and its installed UV lamp These registers are l...

Page 20: ...te low word CLOD 0x0479 ControllerStatusInfo0 Lamp burning Status of Reduced Arc Current dimming Lamp Validation Result Lamp Life Status UV Sensor Status UV Level Rating Relay State CLOD 0x047A Contro...

Page 21: ...Lamp Type Input Register Address 0x0374 This register identifies the specific model of UV lamp that has been attached to the controller This value applies to the last lamp for which the lamp tag was s...

Page 22: ...3 7 3 2 Lamp Rated Operating Time Input Register Address 0x0375 This register displays the rated operating time of the UV lamp in hours This value applies to the last lamp for which the lamp tag was s...

Page 23: ...upport 800 433 5654 Mon Fri 8 am 5 pm EST 23 of 38 3 7 3 3 Controller Operating Time Input Register Address 0x0470 0x0471 These registers display the cumulative operating time of the controller The Co...

Page 24: ...put Register Address 0x0472 0x0473 These registers display the cumulative time that the lamp has been operating The value is specified in units of 15 minute intervals This value applies to the last la...

Page 25: ...ster Address 0x0475 This register displays the number of days of operation that remain for the lamp before it will have been operated for its rated maximum lifetime or longer This value applies to the...

Page 26: ...sured UV Level Absolute Input Register Address 0x0476 0x0477 These registers display the measured UV intensity level reported by the external UV sensor The value is reported in units of 0 01W m2 The v...

Page 27: ...OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 27 of 38 3 7 3 7 Controller Status Information Group 0 Input Register Address 0x0479 This register displays various runtime op...

Page 28: ...Lync WQ UV SECTION 3 Operation L OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 28 of 38...

Page 29: ...Lync WQ UV SECTION 3 Operation L OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 29 of 38...

Page 30: ...OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 30 of 38 3 7 3 8 Controller Status Information Group 1 Input Register Address 0x047A This register displays various runtime op...

Page 31: ...Lync WQ UV SECTION 3 Operation L OMM 006_0A 1 26 2022 Technical Support 800 433 5654 Mon Fri 8 am 5 pm EST 31 of 38...

Page 32: ...feed water meets the feed water requirements within this manual UV sensor fouled or defective Clean or replace UV sensor Low power to lamp Replace controller Lamp intensity does not increase with wate...

Page 33: ...st 5 1 WQ UV Skid Parts Figure 5 1 WQ UV Skid Parts Item Description Part 1 Lync Filter Cartridge 5 Micron LCB 150 P5 68100632 2 Lync Filter Housing LCH 150 68110325 3 Lync UV UV 50 GPM D LD050AGAA 68...

Page 34: ...Parts Item Description Part 1 Pressure relief assembly includes cap spring and stem 68100621 2 Lid with brass reinforced gauge port 68100616 3 O ring for lid closure EPDM 68100619 4 Adapter 2 PVC 681...

Page 35: ...ure 5 3 Lync UV Parts Item Description Part 1 Flow Switch with Cable for All Systems 68108175 2 UV Sensor Accessory Kit with Cable for Controller 68108148 3 Plug for UV Sensor Port when Sensor is Not...

Page 36: ...or replaced with a new cartridge at the maximum of 30 psi differential pressure 1 Close inlet valve 2 Close outlet valve 3 Press pressure relief button to relieve pressure 4 Remove lid after pressure...

Page 37: ...ocedure WARNING If any water comes in contact with any electrical components dry the components immediately DO NOT energize a system with damp or wet electrical components 1 Disconnect the power suppl...

Page 38: ...ower to the system 16 Verify new bulb has 365 days remaining on the controller screen Immediately after replacing the UV lamp and or quartz sleeve follow the disinfection procedure below 6 3 Disinfect...

Reviews: