1.  General information
2.  Design and function
3.  Safety
4.  Transport, packaging and storage
5.  Commissioning, operation
6.  Faults
7.  Maintenance and cleaning
8.  Dismounting, return and disposal
9.  Specifications

Pressure switch, high adjustability of switch differential

For the process industry

Model PSM-700

WIKA operating instructions, pressure switch, high adjustability of switch differential, model PSM-700

1.  General information

 

The mechanical pressure switch described in the operating instructions has been 

designed and manufactured using state-of-the-art technology. All components 

are subject to stringent quality and environmental criteria during production. Our 

management systems are certified to ISO 9001.

 

These operating instructions contain important information on handling the instru-

ment. Working safely requires that all safety instructions and work instructions are 

observed.

 

Observe the relevant local accident prevention regulations and general safety 

regulations for the instrument’s range of use.

 

The operating instructions are part of the product and must be kept in the immedi-

ate vicinity of the pressure measuring instrument and readily accessible to skilled 

personnel at any time.

 

Skilled personnel must have carefully read and understood the operating instruc-

tions prior to beginning any work.

 

The manufacturer‘s liability is void in the case of any damage caused by using 

the product contrary to its intended use, non-compliance with these operating 

instructions, assignment of insufficiently qualified skilled personnel or unauthorised 

modifications to the instrument.

 

The general terms and conditions contained in the sales documentation shall apply.

 

Subject to technical modifications.

 

Further information:

- Internet address:

www.wika.de / www.wika.com

- Relevant data sheet:

PV 35.05

Abbreviations, definitions

DPDT Double pole double throw
SPDT Single pole double throw
NC

Normally closed contact type (NC = normally closed)

NO

Normally open contact type (NO = normally open)

COM Common contact
GND Ground connection

2.  Design and function

2.1  Overview

2.2  Description

The pressure element of the model PSM-700 is a resilient bellow from stainless steel 

which works against a spring mechanism with an adjustable pre-load force. On the 

spring mechanism there is a contact arm for actuating the switch contact. The switch 

is actuated as soon as the force generated by the pressure in the pressure element is 

greater than the set pre-load force.

2.3  Terms used

Maximum operating pressure

The maximum static pressure with which the instrument can be used without changing 

the assured performance data.

Reset point

The pressure value at which the switch returns to the starting position. Mathematically, 

the pressure value for the reset point is equal to the pressure value of the switch point 

minus the switch differential on rising pressure. On falling pressure, the pressure value 

for the reset point is equal to the pressure value of the switch point plus the switch 

differential.

Switch differential

The switch differential is the difference between the switch point and the reset point. 

This is also known as the switch hysteresis.

2.4  Scope of delivery

Pressure switch, operating instructions
Cross-check scope of delivery with delivery note.

Switch differential setting (option)

Blow-out device

Electrical connection 1, lateral

Electrical connection 2, top 

(only for DPDT or option)

Removable cover

Switch point setting

Locking plate with screw (tamper proofing)

Switch point setting display

Process connection

Summary of Contents for PSM-700

Page 1: ...ty and environmental criteria during production Our management systems are certified to ISO 9001 These operating instructions contain important information on handling the instru ment Working safely requires that all safety instructions and work instructions are observed Observe the relevant local accident prevention regulations and general safety regulations for the instrument s range of use The ...

Page 2: ... of any damage caused by using the product contrary to its intended use non compliance with these operating instructions assignment of insufficiently qualified skilled personnel or unauthorised modifications to the instrument The general terms and conditions contained in the sales documentation shall apply Subject to technical modifications Further information Internet address www wika de www wika...

Page 3: ... that lie within its technical performance limits e g max ambient temperature material compatibility For performance limits see chapter 9 Specifications This instrument is not permitted to be used in hazardous areas The instrument has been designed and built solely for the intended use described here and may only be used accordingly The manufacturer shall not be liable for claims of any type based...

Page 4: ...its original packaging in a location that fulfils the conditions listed above 5 Commissioning operation Before installation commissioning and operation ensure that the appropriate instru ment has been selected in terms of design and specific measuring conditions Tools 6 5 mm socket wrench cross head screwdreiver 20 mm open ended spanner 6 mm Allen key 5 1 Requirements at the measuring point The pr...

Page 5: ...operty and the environment Should a failure occur aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument For these media in addition to all standard regulations the appropriate existing codes or regulations must also be followed 1 Remove 4 screws and open cover 2 Remove the cover s of the electrical connection s 3 Fit suitable cable glan...

Page 6: ...ch point Carry out switch point setting with matching test assembly see chapter 5 2 Vibrations Decouple the instrument mechanically Short circuit Moisture in the instrument Only use in ambient conditions for which the ingress protection is suitable Contact chatter repeated short duration opening and closing Vibrations Decouple the instrument mechanically 5 5 Setting of the switch differential opti...

Page 7: ...elow 1 Before cleaning correctly disconnect the instrument from the pressure supply and switch the load circuit off 2 Use the requisite protective equipment 3 Clean the instrument with a moist cloth Electrical connections must not come into contact with moisture CAUTION Damage to the instrument Improper cleaning may lead to damage to the instrument Do not use any aggressive cleaning agents Do not ...

Page 8: ...s I per IEC 61010 1 2010 case grounded with protective conductor overvoltage category II pollution degree 2 Contact version Switch type Electrical rating AC Electrical rating DC Resistive load Inductive load Resistive load Inductive load 125 V 250 V 125 V 250 V 30 V 125 V 250 V 30 V 125 V 250 V Switch differential Fixed SPDT 10 A 10 A 10 A 10 A 2 A 0 4 A 0 2 A 1 A 0 03 A 0 02 A DPDT 15 A 10 A 10 A...

Reviews: