YT-2
Infrared Thermometer

User Manual And Technical Instruction

Please read the user manual carefully and follow the instructions 

before use. For date of manufacture, please refer to the packing.

13. Warranty card

Warranty card

Product name:

 Infrared thermometer

Model:

YT-2

Revision date:03Dec19  

12. Product warranties

1. The thermometer is initially calibrated at the time 
of manufacture.If at any time
Youquestion 

the 

accuracy 

of 

temperature 

measurements,please contact customer service.
2.This  product  provides  free  maintenance  service 
for one year. If you need to provide circuit diagram, 
necessary materials and maintenance of electrical 
circuit  for  any  problem,  please  contact  the 
manufacturer.
3. No free warranty service will be provided for the 
faults caused by the following users:
1) failure caused by unauthorized disassembly and 
modification of products.
2) failure caused by product drop.
3) failure caused by failure to operate according to 
the instructions.
4)failure 

caused 

by 

lack 

of 

reasonable 

maintenance.
5) damage caused by external force.
4.This  product  will  not  cause  allergic  reaction  and 
harm to human body during normal use.

14. Electromagnetic compatibility
 information

Table 1-Compliance information for Emission test

Emission test 

Compliance 

Group 1

RF emissions 

CISPR 11

RF emissions 

CISPR 11

Class B

Immunity
test

Compliance
level

Electrostatic
disc harge
(ESD)
IEC 61000-4-2

Power frequency(50Hz)
magnetic field
IEC 61000-4-8

±8 kV contact
±15 kV air

30A/m
50Hz or 60Hz

Radiated RF EM fields
IEC 61000-4-3

10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz

Table  3-Test  specifications  for  ENCLOSURE 
PORT IMMUNITY to RF wireless communications
 equipment

         Portable RF communications equipment
(including peripherals such as antenna cables and 
external antennas) should be used no closer than 
30 cm (12 inches) to any part ,including cables 
specified by the manufacturer. Otherwise, 
degradation of the performance of this equipment 
could result.

        This equipment might not offer adequate
protection to radio-frequency communication
services. The user might need to take mitigation 
measures, such as relocating or re-orienting the 
equipment.

         When the instrument is in use,never put it near 
other instruments or stack it on other
instrument.if you have to put it near other 
i n s t r u m e n t s o r s t a c k i t o n o t h e r
instruments,please inspect and verify if the 
instrument could run normally.

There is the potentia risk of radio frequency 
interference between the device and other
devices. If there is, please find out the problems 
and take the following measures:
(1) Turn off the device, and turn on again.
(2) Change the direction of the device.
(3)Keep the product away from the interferential 
devices.

Table 2-Compliance information for Immunity test

 Test
frequency
 (MHz)

Band

a)

(MHz)

Service

a)

Modulation

b)

704-787

450

385

800-960

380-390

430-470

TETRA 400

GMRS 460,

 FRS 460 

LTE Band 

13,17

GSM 800/900,

TETRA 800,

iDEN 820,

CDMA 850,

LTE Band 5

710

745

780

810

870

930

Pulse

modulation

b)

 18 Hz

FM

±5 kHz 

deviation

 1 kHz sine

Pulse

modulation

b)

 217 Hz

Pulse

modulation

b)

 18 Hz

 Test
frequency
 (MHz)

Maximum
 power
 (MHz)

Distance

 (m)

 IMMUNITY

TEST LEVEL

 (V/m)

450

385

5240

5500

5785

710

745

780

810

870

930

1720

1845

1970

2450

1700-1990

2400-2570

5100-5800

GSM 1800;

TETRA 1900;

GSM 1900;

DECT;

LTE Band 1,3,

4,25; UMTS

Bluetooth,

WLAN,

802.11 b/g/n

RFID 2450,

LTE Band 7

WLAN 802.11

a/n

5240

5500

5785

1720

1845

1970

2450

Pulse

modulation

b)

217 Hz

Pulse

modulation

b)

217 Hz

Pulse

modulation

b)

217 Hz

1,8

2

0,3

0,3

27

28

0,2

0,3

 9

2

0,3

28

2

0,3

 28

 0,2

0,3

9

2

0,3

 28

NOTE If necesary to achieve the IMMUNITY TEST 
LEVEL, the distance between the transmitting 
antenna and the ME EQUIPMENT or ME SYSTEM 
may be reduced to 1 m. The 1 m test distance is 
permitted by IEC61000-4-3.

a) For some services, only the uplink frequencies are
 included.
b) The carrier shall be modulated using a 50% duty 
cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse 
modulation at 18 Hz may be used because while it 
does not represent actual modulation, it would be 
worst case.

1.Battery replacement: When "    " symbol is 
displayed, Replace with 2 new AAA batteries 
,slide open the battery cover and remove old 
batteries.Replace the batteries being sure to align 
properly as indicated inside the battery 
compartment. 
2.Remove the battery from the product if it is not 
required for extended periods of time in order to 
avoid damage to the thermometer resulting from a 
leaking battery.
3.To protect the environment,dispose of empty 
batteries at appropriate collection sites according 
to national or local regulations.

European Representative: Shanghai International 

Holding Corp.GmbH(Europe) 

Address: Eiffestrasse 80,20537 Hamburg Germany 

VAT: DE 166 892 350

http://www.yuwell.com

Manufacturer:JIANGSU YUYUE MEDICAL 

EQUIPMENT & SUPPLY  CO.,LTD.

Manufacturing Address:No.1 Baisheng Road,Development 

Zone,Danyang,Jiangsu PR 212300 CHINA.

1

7

30206-0A

Summary of Contents for YT-2

Page 1: ...YT 2 Infrared Thermometer User Manual And Technical Instruction Please read the user manual carefully and follow the instructions before use For date of manufacture please refer to the packing ...

Page 2: ...on None 2 Device structure and composition 1 Device use and applicant scope Appendix Instructions 2 AAA Alkaline Batteries Device main structure and composition The thermometer consists of a housing a sensor a display and a circuit board LCD area temperature probe ranging probe power on measurement button battery cover LCD display area battery reminder distance unit temperature unit distance promp...

Page 3: ... the fire to avoid the battery explosion Do not use the battery when it leaks or molds when discarding batteries or this product it should be following local regulations to avoid contamination 7 This product contains sensitive electronic components should avoid strong electromagnetic interference 8 Do not modify this equipment without the authorization of the manufacturer which may lead to measure...

Page 4: ...pose of it in the household waste but take it to appropriate local collection points Setting measure unit In the shutdown state long press the staring measuring button for 8 12 seconds to enter the temperature unit conversion state and then short press the staring measuring button to select and temperature unit then long press the staring measuring button final confirmation battery compartment and...

Page 5: ...ce is less than about 5 cm the device will automatically measure after the measurement it will vibrate and have di sound and measurement result will be displayed When temperature is 37 6 99 7 or higher the device will make quick continuous three time di di di alarm sound when the measured result is beyond device measurement range 32 43 89 6 109 4 thermometer will make quick continuous three time d...

Page 6: ...ing methods otherwise it may cause lens surface of senor abrasion or machine malfunction Disinfection of the product and its head Gently wipe the surface of the product and the head used for measurement with a cotton swab dipped in a little medical alcohol 70 it can be used until the alcohol has completely evaporated Precaution for storage This product should be stored in a dust free dry place ple...

Page 7: ...tronics devices according to Directive 2002 96 EC IP22 EC REP Temperature range Humidity range Atmospheric pressure range This side up Fragile Non rainproof Prevent the invasion of solid objects larger than 12mm When the lamp is tilted 15 degrees it can still prevent the invasion of dripping water When the lamp is tilted to 15 degrees from the vertical Angle the dripping water will not cause harmf...

Page 8: ...f flammable anesthetic gases mixed with air or oxygen or nitrous oxide Product size 143 41 42 mm Product weight about 69g not including cell Number of memory groups 10 groups service life 5years The clinical thermometer is a direct mode Battery replacement cycle use the new battery for no less than 3000 measurements 9 Product technical parameters 11 Replacing the batteries The normal body temperat...

Page 9: ...ication of products 2 failure caused by product drop 3 failure caused by failure to operate according to the instructions 4 failure caused by lack of reasonable maintenance 5 damage caused by external force 4 This product will not cause allergic reaction and harm to human body during normal use 1 Battery replacement When symbol is displayed Replace with 2 new AAA batteries slide open the battery c...

Page 10: ...tection to radio frequency communication services The user might need to take mitigation measures such as relocating or re orienting the equipment When the instrument is in use never put it near other instruments or stack it on other instrument if you have to put it near other i n s t r u m e n t s o r s t a c k i t o n o t h e r instruments please inspect and verify if the instrument could run no...

Page 11: ...e 3 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Table 2 Compliance information for Immunity test Test frequency MHz Banda MHz Servicea Modulationb 704 787 450 385 800 960 380 390 430 470 TETRA 400 GMRS 460 FRS 460 LTE Band 13 17 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 710 745 780 810 870 930 Pulse modulationb 18 Hz FM 5 kHz deviation 1 kHz sin...

Page 12: ...h WLAN 802 11 b g n RFID 2450 LTE Band 7 WLAN 802 11 a n 5240 5500 5785 1720 1845 1970 2450 Pulse modulationb 217 Hz Pulse modulationb 217 Hz Pulse modulationb 217 Hz 1 8 2 0 3 0 3 27 28 0 2 0 3 9 2 0 3 28 2 0 3 28 0 2 0 3 9 2 0 3 28 NOTE If necesary to achieve the IMMUNITY TEST LEVEL the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m The 1 m test...

Page 13: ...ncies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alternative to FM modulation 50 pulse modulation at 18 Hz may be used because while it does not represent actual modulation it would be worst case ...

Page 14: ...tional HoldingCorp GmbH Europe Address Eiffestrasse80 20537HamburgGermany VAT DE166892350 http www yuwell com Manufacturer JIANGSUYUYUEMEDICAL EQUIPMENT SUPPLY CO L TD ManufacturingAddress No 1BaishengRoad Development Zone Danyang JiangsuPR212300CHINA 1730206 0A ...