Fingertip
Pulse Oximeter
USER MANUAL
Ver1.0
General Description
SpO
2
stands for peripheral capillary oxygen saturation. Oxygen saturation is defined as the ratio of oxyhemoglobin (HbO
2
) to the
total concentration of hemoglobin (i.e. Oxyhemo reduced hemoglobin) present in the blood. It is an important physiological
parameter involved in respiration and circulation. The Pulse Oximeter feature herein is small, portable, non-invasive and easy to
use. The user only needs to insert a finger into the chamber to measure his/her SpO
2
level and pulse rate. It has also been proven
to be highly precise and reliable in clinical tests.
Measurement Principle
Oxygenated blood absorbs light preferentially at 905nm (near infrared light), whereas deoxygenated blood absorbs light
preferentially at 660nm (red light). A pulse oximeter works by passing a beam of red and infrared light through a pulsating capillary
bed and then measure the amount of red and infrared light emerging from the tissues via a sensor. To improve accuracy, the 430P
uses a proprietary algorithm to collect data from pulsatile arterial blood and excludes noise from the tissues. The relative absorption
of light by oxyhemoglobin (HbO) and deoxyhemoglobin is then calculated according to the Beer-Lambert's law and a quantitative
measurement of the users’ oxyhemoglobin status i.e. Oxygen saturation level (SpO
2
) is derived.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Sensor
Precautions For Use
1.
Please read the manual carefully before use.
2.
Do not use the fingertip pulse oximeter in an MRI or CT environment.
3.
Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms.
4.
This device is not for continuous monitoring.
5.
Do not use the fingertip pulse oximeter in an explosive environment.
6.
In order to ensure proper sensor alignment and skin integrity, the maximum application time at a single site for the device
should be less than half an hour.
7.
Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not
intended for sterilization.
8.
Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components,
including batteries.
9.
This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or
systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or
strength of a source might disrupt the performance of this device.
10.
Portable and mobile RF communications equipment can affect medical electrical equipment.
11.
Do not disassemble, repair or modify the equipment.
12.
It may be unsafe to
:
—
use accessories, detachable parts and materials not described in the instructions for use
—
interconnect this equipment with other equipment not described in the instructions for use
—
disassemble, repair or modify the equipment
13.
The medical silicone and ABS plastic enclosure which contact the
user’s skin when the device is used have been assessed
by and passed the ISO10993-5 Tests for in vitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type
hypersensitivity.
14.
The fingertip pulse oximeter is dependent on blood flow to obtain an accurate SpO
2
measurement. Verify that nothing is
hindering your blood flow before taking your SpO
2
readings.
15.
This device is not intended to diagnose or treat any medical condition or disease. It is intended for non-medical
use by healthy people to monitor their pulse rate and blood oxygen levels. It is for sports and/or aviation use.
People who need SpO
2
and pulse rate measurements because of a medical condition should not use this pulse
oximeter and should consult with their physician.
Inaccurate measurements may be caused by
1.
Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin).
2.
Intravascular dyes such as indocyanine green or methylene blue.
3.
High ambient light. Shield the sensor area if necessary.
4.
Excessive user movement.
5.
High-frequency electrosurgical interference and defibrillators.
6.
Venous pulsations.
7.
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line.
8.
The user has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
9.
The user is in cardiac arrest or is in shock.
10.
Fingernail polish or artificial fingernails.
11.
Weak pulse quality (low blood perfusion).
12.
Low hemoglobin.
Features
1.
Easy to operate. Measure your SpO
2
non-invasively
2.
Small, light and portable
3.
OLED screen displays SpO
2
, Pulse Rate, Pulse Bar and Plethysmograph.
4.
Level 1-10 adjustable brightness.
5.
6 display modes.
6.
2pcs AAA-size alkaline batteries; low-battery indicator and low power consumption.
7.
When no or low signal is detected, t
he screen will display ‘Finger Out’ and the device will power off automatically in about 8
seconds.
Intended Use
The 430P Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial
hemoglobin (SpO
2
) and pulse rate of children (>2 years old) and adult . It is for sports and/or aviation use.
Recommended age range:
≥
2years old child or adult with small fingers.
Recommended weight range: 33-99 lbs (15-45kg)
Finger size range:0.2
– 0.6 inches (5-16mm)
Operation Instructions
1.
Install two AAA batteries according to the Battery Installation instructions.
2.
Place one of your fingers into the finger chamber of the pulse oximeter.
3.
Press the on/off button on the front panel to turn the pulse oximeter on.
4.
Keep your hand and finger still for the reading. It is recommended that you do not move your body while taking a reading.
5.
Read the data from the display screen.
6.
Press (<1 sec) the button again to toggle between six display modes (see below).
7.
Press and hold (>1 sec) the on/off button to adjust the brightness of the display. There are 10 levels of brightness, with 10
being the brightest. The default level is four.
After turning the Oximeter on, each time you press the on/off button, the Oximeter will switch to another display mode.
There are 6 display modes shown as follows:
1.
2.
3.
4.
5.
6.
Tips to getting a good reading
1.
Make sure that your finger is inserted deep into the chamber so that the fingertip is placed directly in between the LED
sensor and the LED light source.
2.
Avoid making any body movement, especially your finger while taking measurement.
3.
Long fingernails may obstruct the light sensor and prevent accurate measurement. Please keep the fingernails short while
using the device.
4.
Excessive ambient infra-red light, especially in an overly bright lit room, can interfere with the sensor, preventing an accurate
measurement.
5.
Poor blood circulation can affect oximeter readings. Warm your hands and fingers before taking your measurements. Note
that the pulse oximeter is measuring your SpO
2
and PR based on your blood flow. If the blood flow in your finger drops
below a certain level, the pulse oximeter will not be able to get a reading.
6.
Some people with medical conditions such as anemia, hypotension and hypothermia may experience an inaccurate reading
during use. In such case, we suggest that you consult a physician.
Front Panel
The PR (Pulse Rate) Bar Graph and Plethysmograph (SpO2% Waveform) are useful features that can be used to determine the
reliability of a reading. If the height of the pulse bar is less than 30%, this indicates signal inadequacy and the displayed SpO
2
or
pulse rate value is potentially incorrect. Adjust your finger so that it is directly between the LED lights and sensor.
The plethysmograph indicates the amount of blood flow detected by the pulse oximeter. Each wave in the plethysmograph
corresponds to a heartbeat and the wave amplitude corresponds to the amount of blood detected by the pulse oximeter. The pulse
oximeter is optimized when the height of the wave amplitude is consistent throughout as shown in the figure above. That is when
you should take the reading. If the wave amplitude is very small, warm your hands to increase blood flow and retake your
measurement. If the waveform is not consistent, try to keep your hand and body as still as possible.
Package Content
1.
One pulse oximeter
2.
One silicone cover
3.
One lanyard
4.
Two AAA batteries
5.
One instruction manual
6.
One carrying bag/pouch
Battery Installation
1.
Open the battery door cover as shown in the figure.
2.
Install two AAA batteries into the battery compartment in opposite polarity by
matching the plus (+) and minus (-) signs in the compartment. Note: Incorrectly
installed batteries may damage the device.
3.
Close the battery door cover.
Note:
Please remove the batteries if the pulse oximeter will not be used for a long period of
time.
Please replace the batteries when the power indicator starts flickering.
Using the Lanyard
1.
Thread the thin end of the lanyard through the lanyard hole on the device.
2.
Thread the thick end of the lanyard through the thin loop (threaded in step 1) and pull to
tighten.
Warnings!
Keep the oximeter away from young children. Small items such as the battery door,
battery, and lanyard are choking hazards.
Do not hang the lanyard from the
device’s electrical wire.
Please note that the lanyard tied to the oximeter may cause strangulation due to its
length.
Maintenance and Storage
1.
Replace the batteries when the low power indicator appears.
2.
Clean the finger chamber and surface of the oximeter before use.
3.
Remove the batteries if you are not planning to use the oximeter for a long period of time.
4.
Store the product in a cool and dry place, Extreme moisture may damage the oximeter or affect its lifespan.
5.
Dispose battery properly; follow any applicable local battery disposal laws.
Cleaning the fingertip pulse oximeter
It is recommended to clean the oximeter before and after use. To clean, use a soft cloth lightly dampened with water to wipe the
finger chamber and the surface of the oximeter. Allow the oximeter to dry thoroughly before use.
CAUTION:
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device.
Disinfecting the fingertip pulse oximeter
The oximeter can be disinfected as needed. To disinfect, use a soft cloth lightly dampened with 70% isopropyl alcohol. Other
recommended disinfectants include: 70% ethanol or glutaraldehyde-type 2% liquid disinfectants.
CAUTION:
Do not use EtO (Ethylene oxide) or formaldehyde for disinfection.
The lifespan of the pulse oximeter is estimated to be five years if 15 measurements are taken per day and each
measurement takes 10 minutes. Stop using and contact local service center if one of the following incidents occurs:
An error in the
Possible Problems and Solutions
is displayed on the screen.
The oximeter cannot be powered on even though new batteries are installed.
There is a crack on the oximeter or damage to the display resulting in unidentifiable readings.
The spring is defective.
The button is unresponsive.
The pulse oximeter is factory calibrated. You do not need to calibrate it again during its lifespan.
Specifications
1. Display Type
OLED display
2. SpO
2
Display range: 0%~100%
Measurement range: 70%~100%
Accuracy: 70%~100%: ±2%; 0%~69% no definition
Resolution: 1%
3. Pulse Rate
Display range: 30bpm~250bpm
Measurement range: 30bpm~250bpm
Accuracy: 30bpm~99bpm, ±2bpm; 100bpm~250bpm, ±2%
Resolution: 1bpm
430P