background image

Fingertip   

Pulse Oximeter

 

USER MANUAL 

Ver1.1 

General Description 

SpO

2

 stands for peripheral capillary oxygen saturation. Oxygen saturation is defined as the ratio of oxyhemoglobin (HbO

2

) to the 

total concentration of hemoglobin (i.e. Oxyhemo reduced hemoglobin) present in the blood. It is an important physiological 
parameter involved in respiration and circulation. The Pulse Oximeter feature herein is small, portable, non-invasive and easy to 
use. The user only needs to insert a finger into the chamber to measure his/her SpO

2

 level and pulse rate. It has also been proven 

to be highly precise and reliable in clinical tests. 

Measurement Principle 

Oxygenated blood absorbs light preferentially at 905nm (near infrared light), whereas deoxygenated blood absorbs light 
preferentially at 660nm (red light). A pulse oximeter works by passing a beam of red and infrared light through a pulsating capillary 
bed and then measure the amount of red and infrared light emerging from the tissues via a sensor. To improve accuracy, the 500E 
uses a proprietary algorithm to collect data from pulsatile arterial blood and excludes noise from the tissues. The relative absorption 
of light by oxyhemoglobin (HbO) and deoxyhemoglobin is then calculated according to the Beer-Lambert's law and a quantitative 
measurement of the users’ oxyhemoglobin status i.e. Oxygen saturation level (SpO

2

) is derived. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Sensor 

Precautions For Use 

Please read the manual carefully before use. 

Do not use the fingertip pulse oximeter in an MRI or CT environment. 

Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms.   

This device is not for continuous monitoring. 

Do not use the fingertip pulse oximeter in an explosive environment. 

In order to ensure proper sensor alignment and skin integrity, the maximum application time at a single site for the device 
should be less than half an hour. 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 
intended for sterilization. 

Follow  local  ordinances  and  recycling  instructions  regarding  disposal  or  recycling  of  the  device  and  device  components, 
including batteries. 

This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or 
systems. However,  because of the proliferation of radio-frequency transmitting equipment and other sources of  electrical 
noise  in  healthcare  and  other  environments,  it  is  possible  that  high  levels  of  such  interference  due  to  close  proximity  or 
strength of a source might disrupt the performance of this device. 

10 

Portable and mobile RF communications equipment can affect medical electrical equipment. 

11 

Do not disassemble, repair or modify the equipment. 

12 

It may be unsafe to

 

use accessories, detachable parts and materials not described in the instructions for use   

interconnect this equipment with other equipment not described in the instructions for use 

disassemble, repair or modify the equipment 

13 

The medical silicone and ABS plastic enclosure which contact the 

user’s skin when the device is used have been assessed 

by  and  passed  the  ISO10993-5  Tests  for  in  vitro  cytotoxicity  and  ISO10993-10  Tests  for  irritation  and  delayed-type 
hypersensitivity. 

14 

The  fingertip pulse oximeter is dependent on blood flow to obtain an accurate SpO

2

 measurement. Verify that  nothing is 

hindering your blood flow before taking your SpO

2

 readings. 

15 

This device is not intended to diagnose or treat any medical condition or disease. It is intended for non-medical 

use by healthy people to monitor their pulse rate and blood oxygen levels. It is for sports and/or aviation use. 

People who need SpO

2

 and pulse rate measurements because of a medical condition should not use this pulse 

oximeter and should consult with their physician.   

Inaccurate measurements may be caused by 

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin). 

Intravascular dyes such as indocyanine green or methylene blue. 

High ambient light. Shield the sensor area if necessary. 

Excessive user movement. 

High-frequency electrosurgical interference and defibrillators. 

Venous pulsations. 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. 

The user has hypotension, severe vasoconstriction, severe anemia, or hypothermia. 

The user is in cardiac arrest or is in shock. 

10 

Fingernail polish or artificial fingernails. 

11 

Weak pulse quality (low blood perfusion). 

12 

Low hemoglobin. 

Product Features 

Easy to operate. Measure your SpO

2

 non-invasively 

Small, light and portable 

OLED screen displays SpO

2

, Pulse Rate, Pulse Bar and Plethysmograph.   

Level 1-10 adjustable brightness. 

6 display modes. 

2pcs AAA-size alkaline batteries; low-battery indicator and low power consumption.   

When no or low signal is detected, t

he screen will display ‘Finger Out’ and the device will power off automatically in about 8 

seconds. 

Intended Use 

The  500E  Fingertip  pulse  oximeter  is  a  portable  non-invasive  device  intended  for  spot-checking  of  oxygen  saturation  of  arterial 
hemoglobin (SpO

2

) and pulse rate of children (>2 years old) and adult. The device is for sports and/or aviation use.     

Operation Instructions 

Install two AAA batteries according to the Battery Installation instructions. 

Place one of your fingers into the finger chamber of the pulse oximeter. 

Press the on/off button on the front panel to turn the pulse oximeter on. 

Keep your hand and finger still for the reading. It is recommended that you do not move your body 
while taking a reading. 

Read the data from the display screen. 

Press (<1 sec) the button again to toggle between six display modes (see below). 

Press and hold (>1 sec) the on/off button to adjust the brightness of the display. There are 10 levels of brightness, with 10 
being the brightest. The default level is four. 

After turning the Oximeter on, each time you press the on/off button, the Oximeter will switch to another display mode. 
There are 6 display modes shown as follows: 

 

1.

  2.

  3.

  4.

  5.

  6.

 

 

Tips to getting a good reading 

1) Make sure that your finger is inserted deep into the chamber so that the fingertip is placed directly in between the LED sensor 

and the LED light source. 

2) Avoid making any body movement, especially your finger while taking measurement. 

3) Long fingernails may obstruct the light sensor and prevent accurate measurement. Please keep the fingernails short while using 
the device. 

4) Excessive ambient infra-red light, especially in an overly bright lit room, can interfere with the sensor, preventing an accurate 
measurement. 

5) Poor blood circulation can affect oximeter readings. Warm your hands and fingers before taking your measurements. Note that 

the pulse oximeter is measuring your SpO

2

 and PR based on your blood flow. If the blood flow in your finger drops below a certain 

level, the pulse oximeter will not be able to get a reading. 

6)  Some  people  with  medical  conditions  such  as  anemia,  hypotension  and  hypothermia  may  experience  an  inaccurate  reading 
during use. In such case, we suggest that you consult a physician. 

Front Panel 

   

     

The Pulse Bar Graph and Plethysmograph (SpO

2

 Waveform) are useful features that can be used to determine the reliability of a 

reading. If the height of the pulse bar is less than 30%, this indicates signal inadequacy and the displayed SpO

2

 or pulse rate value 

is potentially incorrect. Adjust your finger so that it is directly between the LED lights and sensor.   

The plethysmograph indicates the amount of blood flow detected by the pulse oximeter. Each wave in the plethysmograph 
corresponds to a heartbeat and the wave amplitude corresponds to the amount of blood detected by the pulse oximeter. The pulse 
oximeter is optimized when the height of the wave amplitude is consistent throughout as shown in the figure above. That is when 
you should take the reading. If the wave amplitude is very small, warm your hands to increase blood flow and retake your 
measurement. If the waveform is not consistent, try to keep your hand and body as still as possible. 

Product Accessories 

1. 

One pulse oximeter 

2. 

One silicone cover 

3. 

One lanyard 

4. 

Two AAA batteries 

5. 

One instruction manual

 

6. 

One carrying bag/pouch 

Battery Installation 

1. 

open the battery door cover as shown in the figure. 

2. 

Install  two  AAA  batteries  into  the  battery  compartment  in  opposite  polarity  by  matching  the 
plus (+) and minus (-) signs in the compartment. Incorrectly installed batteries may damage 
the  device.  Note:  In  order  to  achieve  compact  design,  battery  springs  are  unconventionally 
located on the same side. Nonetheless, batteries should be installed in opposite polarity.   

3. 

Close the battery door cover. 

Note:   

 

Please remove the batteries if the pulse oximeter will not be used for a long period of time. 

 

Please replace the batteries when the power indicator starts flickering. 

Using the Lanyard 

1. 

Thread the thin end of the lanyard through the lanyard hole on the device. 

2. 

Thread  the  thick  end  of  the  lanyard  through  the  thin  loop  (threaded  in  step  1) 

and pull to tighten. 

Warnings! 

 

Keep  the  oximeter  away  from  young  children.  Small  items  such  as  the  battery 
door, battery, and lanyard are choking hazards. 

 

Do not hang the lanyard from the 

device’s electrical wire. 

 

Please note that the lanyard tied to the oximeter may cause strangulation due to 
its length. 

Maintenance and Storage 

1. 

Replace the batteries when the low power indicator appears. 

2. 

Clean the finger chamber and surface of the oximeter before use. 

3. 

Remove the batteries if you are not planning to use the oximeter for a long period of time. 

4. 

Store the product in a cool and dry place, Extreme moisture may damage the oximeter or affect its lifespan.   

5. 

Dispose battery properly; follow any applicable local battery disposal laws. 

Cleaning the fingertip pulse oximeter 

It is recommended to clean the oximeter before and after use. To clean, use a soft cloth lightly dampened with water to wipe the 
finger chamber and the surface of the oximeter. Allow the oximeter to dry thoroughly before use.   

CAUTION:

 Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device.   

Disinfecting the fingertip pulse oximeter 

The  oximeter  can  be  disinfected  as  needed.  To  disinfect,  use  a  soft  cloth  lightly  dampened  with  70%  isopropyl  alcohol.  Other 
recommended disinfectants include: 70% ethanol or glutaraldehyde-type 2% liquid disinfectants. 

CAUTION: 

Do not use EtO (Ethylene oxide) or formaldehyde for disinfection. 

The  lifespan  of  the  pulse  oximeter  is  estimated  to  be  five  years  if  15  measurements  are  taken  per  day  and  each 
measurement takes 10 minutes. Stop using and contact local service center if one of the following incidents occurs: 

 

An error in the 

Possible Problems and Solutions 

is displayed on the screen. 

 

The oximeter cannot be powered on even though new batteries are installed. 

 

There is a crack on the oximeter or damage to the display resulting in unidentifiable readings. 

 

The spring is defective. 

 

The button is unresponsive. 

The pulse oximeter is factory calibrated. You do not need to calibrate it again during its lifespan. 

Specifications 

1. Display Type 

OLED display 

2. SpO

2

 

Display range: 0%~100% 
Measurement range: 70%~100%   
Accuracy: 70%~100%: ±2%; 0%~69% no definition 
Resolution: 1%   

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to establish 
the SpO

2

 accuracy. The measured arterial hemoglobin saturation value (

SpO

2

) of the sensors is compared to arterial hemoglobin 

oxygen (

SaO

2

) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to 

the CO-oximeter samples measured over the 

SpO

2

 range of 70%~100%. Accuracy data is calculated using the root-mean-squared 

(Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment

Particular requirements for the basic safety and 

essential performance of pulse oximeter equipment for medical use. 
A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the specified calibration curve and 
the PR accuracy. 
The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3. 

3. Pulse Rate 

Display range: 30bpm~250bpm   

500E 

Reviews: