Summary of Contents for AxioObserver D1

Page 1: ......

Page 2: ...Open Program AxioVision Rel 4 8 Shut Down Procedure 1 If used oil gently remove oil from objective using neatly folded dry lens paper 2 Put 10x objective in place and lower objective 3 Close AxioVisi...

Page 3: ...hat channels you want to image ie All Channels 4 Select the colors tab 5 You can activate deactivate channels by right mouse clicking on the channel number Set Sample on Stage and Focus Only use slide...

Page 4: ...each channel 6 Press the button Adding a Scale bar 1 To add a scale bar press the button or Annotations from the main menu and select Scale bar 2 Drag cursor along image to get desired scale bar size...

Page 5: ...hod HAL Lamp Voltage TL Shutter open RL Shutter closed Reflector Located on bottom front of microscope Controls HAL lamp voltage intensity Diaphragm Control Condenser contrast method Condenser apertur...

Page 6: ...r Correction UV Transmission IR Transmission DIC DIC II Phase Contrast Ph1 Ph2 Catalog 441031 0000 000 421351 9970 710 420650 9901 00 4 5 6 Plan NeoFluar 40x Plan Apochromat 63x Plan NeoFluar 5x 1 3 O...

Page 7: ...Plan NeoFluar 100x 1 3 Oil Ph3 100x 1 3 0 2 0 17 Oil ICS Ph3 000000 1031 171...

Page 8: ...sion BP 460 50 3 Filter Set 38 Cy2 GFP Excitation BP 470 40 Beam Splitter FT 495 Emission BP 525 50 4 Filter Set 46 HE YFP Excitation BP 500 25 Beam Splitter FT 515 Emission BP 535 30 5 Filter Set 45...

Page 9: ...Filter set...

Page 10: ......

Page 11: ...https www micro shop zeiss com p us f f...

Page 12: ...g Time You are allowed to sign up for a maximum of 3 hours per day Please make sure you only book time that you need You are responsible for showing up for your time booked A 10 00 fee applies if you...

Page 13: ...load your sample on the microscope stage with the objective lowered to the lowest focus position Oil should never reach the outer rim of the objectives The felt protectors should never become saturate...