Directions for Zeiss Axiovert 200M 

 

START UP 
1.

 

Turn on mercury lamp controller (#1). Wait a few seconds to make certain lamp is 
on. 

2.

 

Turn on the x-y stage controller (#2) 

3.

 

Turn on the camera controller (#3) 

4.

 

Turn on the microscope (#4) (green button on the right side of the microscope. 

5.

 

Login to the computer with your CNET ID and password. Double click on the 
Openlab program on the computer. 

 
IMAGING 
1.

 

Choose the correct stage for slide or dish. Make sure the stage is properly inserted 
(there is a spring in the corner). The stage should be laying level: Don’t have one 
side higher than the other- it should not wobble. If your sample is on a slide, make 
sure the coverslip is facing down. Choose the objective you want by moving it in 
to place or choose the objective on the Microscope palette. The dry objectives are 
10x, 20x, and the first 40x.  If you are using an oil objective (the second 40x, 63x 
or 100x) put 1 drop of oil directly onto the objective. When focusing with the oil 
objective you should have a space of oil between the objective and the cover slip. 
The objective and cover slip should NOT be touching. 

 
 

 

 

slide 

 

Coverslip 

 

 

 
2.

 

Check the fluorescent palette in the program: if the filter you need is not in place, 
use the red- handled Zeiss tool and unscrew the screw that holds the reflector 
turret in place with 2 turns. (DON’T unscrew until the screw falls out). Gently 
push the reflector turret out and exchange filter cubes. (Do not touch the actual 
filters with your fingers). The cubes should all be level to each other if properly 
inserted in the turret. Carefully replace the reflector turret (it should be smooth 
and easy to insert- do not force it! If you are forcing it, then you are inserting it 
incorrectly, and it will get jammed!) Use the Zeiss tool to lock in the reflector 
turret after changing your filter. Then change the label for the filter in the reflector 
pallet in the program.  

3.

 

If possible start by imaging your sample using 

transmitted light

 to prevent 

bleaching. Click on Transmitted Light or the dic (#5 position) in the Zeiss 
reflector palette and focus using the focusing dial on the right side of the 
microscope. 

(Do not open the fluorescent shutter at any time while on 

transmitted light!) 

Use Koehler illumination to focus the condenser for 

transmitted light every time you switch to a new objective. For transmitted light, 
choose the camera shutter that sends 50% of the light to the camera and 50% of 
the light to the ocular. 

a.

 

Koehler Illumination: After focusing on your sample with transmitted 
light, close the diaphragm, and use the condenser-focusing knob on the 
right side of the transmitted lamp to focus the condenser unit so that the 
diaphragm is sharply focused. You should now see your sample and the 

Oil objective 

oil 