Zeiss LSM800-6 vertical-2 + Airyscan confocal microscope  

   

2019-03-15                   

 

I.

 

System description: 

Inverted (vertical) confocal microscope with Airyscan module 

Motorized stage for multiple position imaging  

Objectives:  

  Objective Plan-Apochromat 10x / NA 0.45 Air (WD=2.1mm)

 

  Objective Plan-Apochromat 20x / NA 0.8 Air

 

  Objective W-Plan-Apochromat 40x / NA 1.0 water dipping DIC VIS-IR

 

  Objective C-Apochromat 63x / NA 1.2 W Korr

 

 

 

Laser lines: 405, 488, 561, 640 nm 
 

Detectors: 

 

2 PMT detectors with free choice of the spectral range. (PMT 1: tunable short pass and various 
bandpass filters in front of the PMT, PMT 2: tunable long pass and various bandpass filters in 
front of detector) 

 

1 GaASP PMT for the Airyscan detector that can function as a normal detector in confocal mode 
(detection range adjustable by a tunable short pass and a tunable long pass filter, no bandpass 
filters) 

 

An ESID detector for transmitted light. 

Simultaneous detection of only two dyes possible, more than two have to be acquired sequentially. 

 

II.

 

Starting up: 

 

1.

 

Switch on the multiple socket (1). Provides power to microscope controllers 

2.

 

Switch on the system(2) and Components (3) 

3.

 

Switch on the computer (4) 

4.

 

Login: BIF-User (Useme!11) 

5.

 

Open 

ZEN Blue 

Software and press “Start System” (“Image Processing” is for offline analysis of 

your images) 

6.

 

If Pop-up appears about stage calibration, press “calibrate now” 

 
 