background image

                                                                                                                                                     
                                                                                                                                                    Issue date

2020/3/16 

1

 

 

Surgical Microscope 

OMS2350

 

   

User Manual 

 

The  CE  mark  indicates  that  the  product  complies  with  the  European  medical 

device regulation (EU) 2017/745. 

 

  Zumax Medical Co., Ltd. 

5 Zhiying Street, Suzhou New District, Suzhou, Jiangsu, China 215129 

Phone

+86 512 66650502 

Fax

    +86 512 66909655 

http://www.zumaxmedical.com 
E-mail

sales@zumaxmedical.com 

                                 

MedNet EC-REP GmbH 

Borkstrasse 10, 48163 Münster, Germany 

 

 

Summary of Contents for OMS2350

Page 1: ...omplies with the European medical device regulation EU 2017 745 Zumax Medical Co Ltd 5 Zhiying Street Suzhou New District Suzhou Jiangsu China 215129 Phone 86 512 66650502 Fax 86 512 66909655 http www zumaxmedical com E mail sales zumaxmedical com MedNet EC REP GmbH Borkstrasse 10 48163 Mü nster Germany ...

Page 2: ...rument is well grounded 3 Please pay attention to all the rated values of the electrical connecting terminal 4 Only use fuse according to the specifications and rated values stipulated by our product 5 Use the power cable supplied with this instrument only 6 Do not touch the surface of the lens and prism with hand or hard objects 7 Turn off the main power first before replacing the illumination bu...

Page 3: ...MENT The black dot is the video interface The blue dot is the LED power interface Operating instruction s Warning Caution Protective earth ground light dimmer Increase the brightness clockwise Reduce the brightness counterclockwise ESD Balance adjustment ...

Page 4: ...e use 12 4 2 1Adjusting the balance of second arm 12 4 2 2 Adjusting the microscope 12 4 3 Inspection before use 13 4 4 Using the microscope 14 4 5 Movement and Storage after use 14 5 Maintenance 16 5 1Replacing the consumable parts 16 5 1 1Replacing the LED 16 5 1 2 Replacing the fuse 16 5 1 3 Replacing the power supply cords 16 5 2 Cleaning and Sterilization 16 5 2 1 Cleaning the surface of equi...

Page 5: ...tically adjusting the exposure fatigue is minimized even after prolonged use Six step magnification for the main microscope the optical system assures a perfectly clear images Cool illumination by LED lamp Natural colors Hanging arm design so the arm can be moved randomly A special switch has been fitted into the Second Arm The lamp will keep on when the arm is working in its normal range in parki...

Page 6: ...Issue date 2020 3 16 2 2 Main components Fig 2 General assembly of microscope ...

Page 7: ...Issue date 2020 3 16 3 Fig 3 Back side view 27 Fig 4 Integrated full HD camera 28 25 12 13 14 ...

Page 8: ...xation screw Use this screw to adjust the damping of the horizontal movement of the first arm 6 Main power socket fuse socket Pull out the power table right side is fuse socket fuse specification T1 25 A H250 V for 220VAC T2 5 A H250 V for 110VAC 7 Main Power Switch To turn on or turn off the m1ain power of the microscope 8 Foot Wheel Lock the brakes to fix the microscope in place 9 Star handler f...

Page 9: ...ing using this nut 17 Star handler fixation screw Use this handle to adjust the damping of the microscope s hanging axis 18 Star handler fixation screw Use this handle to adjust the damping of the microscope s pitch 19 Locking pin Keep the microscope from falling when assembling and disassembling accessories 20 Fixation screw of main microscope Used to fix the main eyepiece tube 21 Manipulating ha...

Page 10: ...ns as shown in the figure3 Press the left button will take a picture press the right button will record video Interface details on the arm as shown in the figure4 28 Interface details on the arm LAN HDMI SD Card USB Debug Optional 29 Inner Camera Remote Control Kit Optional The details see instruction of inner camera remote control kit ...

Page 11: ...date 2020 3 16 7 3 Assembly The microscope is packed in one package Please open the package and take out all parts and assemble them according to the following procedures Fig 6 Assembly drawing of microscope ...

Page 12: ... Finally fasten the socket screw firmly with an 8mm allen key Fig 7 Assembly of base 3 2 Assembly of the first arm Take out the first arm please pay attention the star handler fixation screw 3 must be tightened in advance insert the hole to the axis then use jump ring pliers to fix the two axis with jump ring in the circle groove put on the cover Caution The Second Arm use a spring to keep the bal...

Page 13: ...the hole of second arm from the bottom to up tighten the locking pin 19 make the locking pin 19 into the axis groove tighten the fixation nut 4 finally Fig 8 Assembly drawing of 120º coupling 3 4 Assembly of microscope head Take out the microscope head connect it with the hanging axis Use the method as same as 3 3 to assemble the microscope head to the 120º coupling 3 5 Assembly of 180ºinclinable ...

Page 14: ...lied with your equipment then tighten the screw According your needs select manipulating or T type handle 3 7 Assembly of the LED power cable The LED power cable is fixed on the first and second arm Plug LED power cable into the joint 12 Caution The terminals of LED power cable must be plugged in the right holes of the joint 3 8 How to connect the power cable Take out the power cable and plug it i...

Page 15: ...cting switch 23 to the 110V block and change the fuse to type T2 5 A H250 V as supplied If you will operate the equipment at 220V please switch the selecting switch 23 to the 220V block and change the fuse to type T1 25 A H250 V as supplied Check the grounding of power supply Make sure the equipment has a good ground wire connection Please use the power cable supplied with this equipment Please ma...

Page 16: ...s adjust the eyepiece to 0 Operators using his her corrective glasses adjust the eyepiece to 0 Ametropes operators who need corrective glasses who know their refractive powers and perform surgery without wearing their glasses adjust the eyepiece to his her own eyesight Ametropes operators who need corrective glasses who do not know their refractive powers and perform surgery without wearing their ...

Page 17: ...witch on and off The LED turns off automatically when the microscope is in parking position The LED power cable has been connected Set balance Ground must be level so the column will be level The damping of first arm second arm 120º connecting arm optional accessories and the microscope has already been adjusted Microscope Adjusting the magnification Magnifying knob is 13 working correctly Eyepiec...

Page 18: ...uests of Inspection Form Turn on the power switch Move the second arm up and down to the right working position Adjust the illumination Select the filter which will be used Move the microscope lens into the operation area then adjust it to a suitable position Adjust the magnification Move the operation microscope adjust rough focusing through the eyepieces then adjust fine focus Move the second ar...

Page 19: ...kes of the wheels While moving the equipment hold the movement handles and make the equipment move slowly and carefully to avoid falling and bumping When the storage place has been reached lock the brakes of the wheels Put on the dust cover ...

Page 20: ...with a small screwdriver Take out the melted fuse Insert a new fuse and insert the fuse socket back again Plug in power cable Turn on the main power switch Specification of fuse T1 25 A H250 V for 220VAC T2 5 A H250 V for 110VAC Caution Please use the bulb specially designed for this equipment 5 1 3 Replacing the power supply cords Replacement the power supply cords should by the manufacturer spec...

Page 21: ...with a little bit of distilled water and wash off the remaining stains can be cleaned off with a mixture of 50 ethanol and 50 ether If there is dust on the lens blow the dust off with a dust pen Do not wipe the lens with any corrosive detergent otherwise it may be damaged 5 2 3 Cleaning and disinfection of sterilizable caps Cleaning and Disinfection in Automatic cleaning machine with pure water ac...

Page 22: ... pcs 800 0133 20X Eyepiece 2 pcs 800 0134 12 5X Eyepiece with target reticle 800 0139 Rubber Eye Guard 2 pcs 800 0151 New 120 deg coupling 800 0152 balance arm 800 0166 Canon Digital Camera Adapter 800 0167 Sony Digital Camera Adapter 800 0168 Nikon Digital Camera Adapter 800 0169 Nikon 1J5 Digital Camera Adapter 800 0170 Canon Digital Camcorder Adapter 800 0171 Nex5 Digital camera Adapter 800 017...

Page 23: ...less digital camera Adapter 800 0312 Wall Mount System with plastic covers 800 0324 Ceiling Mount System for new arm 800 0400 Easy360 mobile phone adapter 800 0410 Plastic case for iPhone of Mobile phone adapter 800 0412 Plastic case for Huawei of Mobile phone adapter 800 0414 Plastic case for Samsung of Mobile phone adapter 800 0416 Plastic case for Sony of Mobile phone adapter 800 0418 Plastic c...

Page 24: ...otswitch for Sony camera 800 9374 Wireless control Kit by footswitch for Nikon camera 800 1106 sterilizable caps for friction adjustment knobs 800 1206 sterilizable small caps for friction adjustment knob 800 1303 sterilizable cap for fine focus 800 1304 sterilizable cap for magnification changer 800 1306 sterilizable cap for rotation handle 800 1307 sterilizable cap for T handle 800 1308 steriliz...

Page 25: ...cable is not inserted to the joint correctly Insert it to the joint correctly Microscope not in working area but in parking position Move the second arm to working area The LED has burnt Contact the after sales service department Illumination brightness adjusting button is in low position Adjust the button a higher position Bulb turns on and off during the surgery Something covered the heat elimin...

Page 26: ...000 lx Illumination diameter of light spot mm Φ80 Stand Section including base column first arm and second arm Second Arm Length 600mm Rotational angle 150 Distance 300mm First Arm Length 500mm Rotational angle 360 Height 1750mm Base dimension 610 600mm Electrical specifications Rated voltage AC230V 10 50Hz AC120V 10 60Hz Input voltage 40VA Fuse AC 120V T2 5 A H250V AC 230V T1 25 A H250V Electrica...

Page 27: ...ransducers and cables other than those specified with the exception of transducers and cables sold by the MANUFACTURER of the microscope as replacement parts for internal components may result in increased EMISSIONS or decreased IMMUNITY of the microscope Guidance and manufacturer s declaration electromagnetic emissions The microscope is intended for use in the electromagnetic environment specifie...

Page 28: ... Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in...

Page 29: ...strengths from stationary RF transmitters as determined by a site surveya should be less than the compliance level in all frequency ranges b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption ...

Reviews: