11-38
Programmer’s Guide
SCPI Command Summary
SCPI Device Command Summary
Table 11-19
SENSe (1 of 16)
1. NR1 values of 1 (on) and 0 (off) can be used in place of CHAR values ON and OFF.
SUBSYSTEM COMMANDS
FORM
DESCRIPTION
SENSe[1|2]:ACTIVE?
query
only
Returns true (1) if the channel is
active, false (0) if the channel is not
active. (Only one channel can be
“active” at a time.)
SENSe[1|2]:AVERage:CLEar
command
only
Re-starts the trace averaging
function.
SENSe[1|2]:AVERage:COUNt
<num>
NR1
Specifies a count or weighting factor
for the averaged measurement data.
SENSe[1|2]:AVERage[:STATe]
<ON|OFF>
1
NR1
Turns the trace averaging function
on/off.
SENSe[1|2]:BWIDth
[:RESolution] <num> HZ
NR2
Specifies the bandwidth of the IF
receiver (fine, narrow, medium or
wide) to be used in the measurement
— choose
15 (fine)
250 (narrow)
1200 (medium narrow)
3700 (medium)
4000 (medium wide)
6500 (wide)
SENSe[1|2]:CORRection:
NR1
Turns measurement calibration
function on/off. Uncorrected readings
are used when “off.”
SENSe[1|2]:CORRection:
ANNotation?
query
only
Returns the current calibration
annotation: "C", "C?", “Cx”, or "".
SENSe[1|2]:CORRection:
CAPacitance:CONNector <num>
NR3
Selects a connector compensating
capacitance value. (For use with
structural return loss measurements
on analyzers with Option 100 only.)
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...