Programmer’s Guide
11-43
SCPI Command Summary
SCPI Device Command Summary
Table 11-19
SENSe (6 of 16)
1. Binary parameters accept the values of
1
(on) and
0
(off) in addition to
ON
and
OFF
.
2. For use with multiport test sets only.
SUBSYSTEM COMMANDS
FORM
DESCRIPTION
SENSe[1|2]:CORRection:COLL
ect:ISTate[:AUTO]<ON|OFF>
NR1
Selects the instrument state for
calibration — choose Full Band (
ON
) or
User Defined (
OFF
).
SENSe[1|2]:CORRection:
COLLect:METHod <char>
command
only
Selects the type of calibration — choose
from:
TRAN1
— Transmission response
TRAN2
— Transmission response &
Isolation
TRAN3
— Transmission enhanced
response
REFL3
— Reflection one port
TESTset
— Test Set Calibration
(Enhanced Response/1-Port)
1
TESTset1
— Test Set Calibration
(Enhanced Response/1-Port)
TESTset2
— Test Set Calibration
(2-Port)
VERIFY
— Calibration Check
TWOPort
— Two-port calibration
NONE
— No calibration
SENSe[1|2]:CORRection:COLL
ect:MP:OPEN
<STAN1|STAN2|…|STAN12>
command
only
Measures an open on the port selected
during a test set calibration.
2
SENSe[1|2]:CORRection:COLL
ect:MP:SHORT
<STAN1|STAN2|…|STAN12>
command
only
Measures a short on the port selected
during a test set calibration.
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...