5-33
Programming Commands
CALCulate2 Subsystem
PEXCursion
Sets the peak excursion limit used by the Agilent 86120B to determine valid
laser line peaks.
Syntax
:CALCulate2:PEXCursion{?| {<integer> | MINimum | MAXimum | DEFault}}
<integer>
represents logarithmic units in dB. Valid range is 1 to 30 dB.
Constant
Description
MINimum 1
dB
MAXimum 30
dB
DEFault 15
dB
Attribute
Summary
Non-sequential command
Preset
State: 15 dB
*RST State: 15 dB
SCPI Compliance: instrument specific
Description
A laser line is identified as a valid peak if its amplitude is greater than the peak
excursion plus the amplitudes of the closest local minima on either side of the
peak. This command works in conjunction with the peak threshold setting.
Refer to “PTHReshold” on page 5-34
. Changing the peak excursion limit
causes the instrument to reprocess the current set of data. Refer also to
“Defining Laser-Line Peaks” on page 2-16
.
The query response is the current value. For example, if the current value is
set to 15 dB, the following value is returned:
15
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......