REMOTE CONTROL
193
[SENSe:]VOLTage:DC:NPLCycles
Sets the integration time for DC Voltage measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:VOLT:DC:NPLC 12
Sets the integration time to 12 PLCs for DC Voltage.
[SENSe:]VOLTage:DC:NPLCycles?
Returns the integration time for DC Voltage measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]CURRent:DC:NPLCycles
Sets the integration time for DC Current measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:CURR:DC:NPLC 2
Sets the integration time to 2 PLCs for DC Current.
[SENSe:]CURRent:DC:NPLCycles?
Returns the integration time for DC Current measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]RESistance:NPLCycles
Sets the integration time for 2W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:RES:NPLC MIN
Sets the integration time to 0.025 PLCs for 2W resistance measurements.
[SENSe:]RESistance:NPLCycles?
Returns the integration time for 2W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12