PDL5 User Manual
PDL5-UM-00001 Rev 001
62
[:POWER]:MODe
Syntax
[:POWER]:MODe<space>[ABS|REL|BRM|PDL]
Function
Switches the meter between the following operation modes:
ABS:
absolute power operation mode
REL:
relative power operation mode
BRM:
backreflection operation mode
PDL:
IL avg and PDL operation mode.
Example
MOD PDL
switches the meter into ILavg and PDL mode
[:POWER]:MODe?
Syntax
[:POWER]:MODe?
(note: v1.00 doesn’t return PDL due to a bug)
Function
Returns the current operation mode.
Example
MOD?
returns: ABS.
[:POWER]:READ?
Syntax
[:POWER]:READ?
Function
Triggers and Returns the reading of the meter
Example
READ?
returns
-55.2 (the unit is in BR mode).
[:POWER]:READ:FULL?
Syntax
[:POWER]:READ:FULL?
Function
Returns the current reading (power or BR) of the meter, the current
channel number, detector number and output wavelength (in non-
PDL mode).
Example
READ:FULL?
returns
-55.2, 5, 0, 1310 (the unit is in BR mode).