34
Buffer
— This same digital data is also stored in the buffer as each channel is scanned (press RECALL to access data stored in
the buffer). Note that if you change the digital data format, the format for digital readings already stored in the buffer will also
change.
Remote programming
The following commands are used to set digital limits. For details, see
and the command descriptions that follow the
table.
CALC3:LIM[1]:DIG:MODE <name>
Limit 1 mode; PATTern or MASK.
CALC3:LIM[1]:DIG:PATT <NDN or NRf>
Limit 1 bit-pattern for comparison.
CALC3:LIM[1]:STAT <b>
Enable or disable limit 1 test.
CALC3:LIM2:DIG:MODE <name>
Limit 2 mode; PATTern or MASK.
CALC3:LIM2:DIG:PATT <NDN or NRf>
Limit 2 bit-pattern for comparison.
CALC3:LIM[2]:STAT <b>
Enable or disable limit 2 test.
Digital limit testing while scanning
Input channels for an advanced scan can also be configured to perform digital limit tests.
1. Press SHIFT and then CONFIG.
2. Select “INT: ADVANCED” and press ENTER.
3. Use the left and right edit keys to display the desired digital channel (111, 112, 113, or 114).
4. Press SHIFT and then CH OFF to enable the channel. If, for example, you enable channel 112, the following will be
displayed:
SETUP D:112
NOTE
If the channel was configured as an output channel (SETUP message dimmed), it will
automatically switch to an input when is enabled.
5. Configure limits for scan channel — Press SHIFT and then LIMITS to enter the limits menu. Use the left, right, up, and
down edit keys and ENTER to configure limits.
To configure digital limit 1, select the mode (pattern or mask) and then set the digital pattern by specifying the equivalent
decimal value (000 to 255). Digital limit 2 is configured in a similar manner.
The following menu messages show digital limit 1 set to bit pattern 00001111 (decimal 15) for the pattern mode:
DL1 MOD: PATT
DL1 PAT: 015
If still in the limits menu structure, use the ENTER key to return to the scan channel edit mode.
6. Enable limits for scan channel — Press SHIFT and then ON/OFF to display the state of limits (off or on). Use the up or
down edit key to display “LIMITS: ON” and press ENTER. The HIGH and LOW annunciators will turn on to indicate that
limits is enabled for the channel.
7. To configure another digital channel, repeat steps 3 through 6.
8. When finished setting up scan channels, press ENTER and finish configuring the scan.
Summary of Contents for 7707
Page 55: ...52 Specifications 7707 Multiplexer Digital I O Module...
Page 58: ......