C-2
Common Commands
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Introduction
This appendix includes information on common commands and their script
command equivalents.
Common commands
Command summary
Common commands supported by the Series 2600 SourceMeter are summarized
in
. Although commands are shown in upper-case, common commands
are not case sensitive, and either upper or lower case can be used. Note that
although these commands are essentially the same as those defined by the IEEE-
488.2 standard, the Series 2600 does not strictly adhere to that standard.
Table C-1
Common commands
Mnemonic
Name
Description
*CLS
Clear status
Clears all event registers and Error Queue.
1
*ESE <mask>
Event enable command
Program the Standard Event Enable Register.
1
*ESE?
Event enable query
Read the Standard Event Enable Register.
1
*ESR?
Event status register query
Read/clear the Standard Event Enable Register.
1
*IDN?
Identification query
Returns the manufacturer, model number, serial
number, and firmware revision levels of the unit.
*OPC
Operation complete command
Set the Operation Complete bit in the Standard
Event Register after all pending commands have
completed.
*OPC?
Operation complete query
Places an ASCII “1” into the Output Queue when
all selected device operations have completed.
*RST
Reset command
Returns the SourceMeter to default conditions.
*SRE <mask>
Service request enable
command
Programs the Service Request Enable Register.
1
*SRE?
Service request enable query
Reads the Service Request Enable Register.
1
*STB?
Status byte query
Reads the Status Byte Register.
1
*TRG
Trigger command
Sends a remote trigger to the SourceMeter.
*TST?
Self-test query
Returns a 0.
*WAI
Wait-to-continue command
Waits until all previous commands have
completed.