6-22
Keysight B1542A User’s Guide, Edition 10
PLSDIV TIS Commands
Command Reference
Write String
This command needs the following input parameters in this sequence.
sampling_rate
Sampling rate. Numeric value, 0 (auto), 1 to 20 G [times/s]. It is
difficult to find the optimum value. Set
sampling_rate
=0 for an
easy setup.
sampling_num
Total number of sampling. Numeric value, 0 (auto), 16 to
262144.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivSetSmoothing
This command enables or disables smoothing of the specified oscilloscope channel.
Execution
Conditions
PlsdivSetSampling command must be executed before this command.
Filename
PlsdivSetSmoothing.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameters in this sequence.
ch
Oscilloscope channel. Numeric value, 0 (channel for gate) or
1 (channel for drain).
smooth_stat
Smoothing function on or off. Numeric value, 0 (off) or 1 (on).
smooth_num
Number of measurement points used for smoothing. Numeric
value, 3 to 4001. The value must be less than the
sampling_num
parameter value set to the PlsdivSetSampling command.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Remarks
The smoothing function averages the time-series data to reduce variation caused by
high-frequency component.
Set appropriate
smooth_num
value while monitoring the waveform. Too much
smooth_num
value will lose the waveform edges, which disturbs accurate results.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......