National Instruments Corporation
33
NI 4070/4072 Calibration Procedure
Verifying AC Current
To verify the AC current of the NI 4070/4072, complete the following
steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072
HI SENSE
and
LO
banana plug connectors, and fasten
the connectors on the other end of the cable to the
HI
and
LO
calibrator
binding posts. Figure 4 shows the correct connections. Table 11 lists
the cable connections.
3.
Call
niDMM_reset
to reset the NI 4070/4072 to a known state.
4.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_CURRENT
•
Range
=
0.01
•
Resolution
=
10e–9
5.
Call
niDMM_ConfigureAutoZeroMode
with the following
parameters:
•
Auto Zero Mode
=
NIDMM_VAL_AUTO_ZERO_ON
6.
Call
niDMM_Read
to configure the NI 4070/4072 for a current mode
before applying current.
7.
Output 1 mA at 1 kHz on the calibrator with the current output set
to
NORM
.
8.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 22.
9.
Output 10 mA at 1 kHz on the calibrator.
10. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 22.
11. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_CURRENT
•
Range
=
0.1
•
Resolution
=
100e–9
12. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 22.
13. Output 100 mA at 1 kHz on the calibrator.
14. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 22.