NI 4071 7½-Digit FlexDMM Calibration Procedure
41. Generate –0.5 A on the calibrator with the current output set to
NORM
.
42. Call the niDMM Cal Adjust Gain VI with the following parameters:
43. Remove all connections from the four input banana plug connectors on the NI 4071.
44. Call the niDMM Cal Adjust Offset VI with the following parameters:
45. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
–
0.5
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1e
–
6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
10e
–
6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA