NI 5105 Calibration Procedure
9.
Disconnect or disable all inputs to the device.
10. Self-calibrate the device using the niScope Cal Self Calibrate VI.
.
11. End the session and save the results using the niScope Cal End VI.
You have finished adjusting the NI 5105. Repeat the
section to reverify the
performance of the NI 5105 after adjustments. Use the calibration test limits to verify the
performance of the product after adjustment, and compare the results with the calibration test
limits.
LabVIEW VI
C/C++ Function Call
Call
niScope_CalSelf
Calibrate
with the
following parameters:
session handle:
The
instrument handle from
niScope_CalStart
channelList
:
VI_NULL
option
:
VI_NULL
LabVIEW VI
C/C++ Function Call
Call
niScope_CalEnd
with the
following parameters:
sessionHandle
: The session handle
from
niScope_CalStart
action
:
NISCOPE_VAL_ACTION_STORE
Table 9.
NI 5105 Input Parameters for External Adjustment
Iteration
Range (V)
Input (V)
1
0.05
0.02
2
0.20
0.09
3
1.00
0.425
4
6.00
1
5
30.00
2.5
6
0.05
-0.02
7
0.20
-0.09
8
1.00
-0.425