*CFF
Turns the power correction OFF
“ACK\r\r\n”
*AOB
Modifies the analog output voltage value (*AOB + 8
characters)
Example :
*AOB1.00E+01
The maximum value of the
analog display (2.05 volts) is 10 watts
“ACK\r\r\n”
*AOD
Sets an analog output delay from 1s to 7s for a return to 0V
after an energy measurement. Default 0 (no return to zero
after an energy measurement).
Example :
*AOD2.00E+00
The delay is 2 seconds.
“ACK\r\r\n”
*RST
Resets the device
“”
*SOU
Sets the zero offset
“ACK\r\r\n”
*KPA
Returns the character string «
ACK
»
“ACK\r\r\n”
*PWC
Wavelength correction value (+ 5 characters)
Example :
*PWC01064
selects the wavelength 1064 nm
“ACK\r\r\n”
*TLC
Modifies the trigger level in energy mode (+ 8 characters)
Example :
*TLC2.00E-02
selects a Trig Level of 0.002
Joules
“ACK\r\r\n”
*NAM
Returns the model of the head to the serial port
“UP55N-400W-
H9\r\r\n”
*VER
Returns the name of the software to the serial port
“Version 1.14\r\r\n”
*F01
Returns information about current status
All field are separate by a TAB character “\t”
“Version\t3\tName\t
UP55N-400W-
H9\tWavelength\t10
64\tTrig
Level\t2.500000e-
01\tMax Analog
Output\t2.
00\tMode\t0
Offset\t0\tMUL\t1.0
00\tOFF\t0.
00\tPWCS
tatus
65536\tMinScale\t2
4\r\r\n”