Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1153
User Manual 1178.8520.02 ─ 01
14.11.1
General Status Register Commands
...............................................................................................1153
STATus:PRESet
This command resets the edge detectors and
ENABle
parts of all registers to a defined
value. All
PTRansition
parts are set to
FFFFh
, i.e. all transitions from 0 to 1 are
detected. All
NTRansition
parts are set to 0, i.e. a transition from 1 to 0 in a
CONDition
bit is not detected. The
ENABle
part of the
STATus:OPERation
and
STATus:QUEStionable
registers are set to 0, i.e. all events in these registers are not
passed on.
Usage:
Event
STATus:QUEue[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
This command is identical to the
Usage:
Query only
14.11.2
Reading Out the CONDition Part
For more information on the condition part see
Chapter 13.1.7.2, "Structure of a SCPI
STATus:OPERation:CONDition?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:ACPLimit:CONDition?
<ChannelName>
STATus:QUEStionable:EXTended:CONDition?
<ChannelName>
STATus:QUEStionable:EXTended:INFO:CONDition?
<ChannelName>
STATus:QUEStionable:FREQuency:CONDition?
<ChannelName>
STATus:QUEStionable:LIMit<n>:CONDition?
<ChannelName>
STATus:QUEStionable:LMARgin<n>:CONDition?
<ChannelName>
STATus:QUEStionable:POWer:CONDition?
<ChannelName>
STATus:QUEStionable:TEMPerature:CONDition?
<ChannelName>
STATus:QUEStionable:TIME:CONDition?
<ChannelName>
These commands read out the CONDition section of the status register.
The commands do not delete the contents of the CONDition section.
Suffix:
<n>
.
Using the Status Register