command steps up or down the current according to the step size set in
[:SOURce[<n>]]:CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
•
to set the current for the specified channel.
Return Format
The query returns the current, for example, 1.500.
Examples
:CURR 1.5 /*Sets the current to 1.5 A for the channel currently
selected .*/
:CURR? /*Queries the current of the channel currently selected. The
query returns 1.500.*/
4.9.2
[:SOURce[<n>]]:CURRent[:LEVel]
[:IMMediate]:STEP[:INCRement]
Syntax
[:SOURce
[<
n>]
]:CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
{<
numeric
value>|<def>}
[:SOURce
[<
n>]
]:CURRent[:LEVel][:IMMediate]:STEP[:INCRement]?
[<
def>]
Description
Sets or queries the step size of current for the specified channel.
Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
<numeric value>
Real
0 to the maximum current
value of the specified channel
Refer to
<def>
Discrete
{DEFault}
-
Remarks
•
If [:SOURce[<n>]] or [<n>] is omitted, the command sets the corresponding
parameter of the channel currently selected.
•
<numeric value> is the step size specified. DEFault is the default value. The
default values of <numeric value> are as shown in the table below.
Channel
Default
DP932A
Standard
CH1
0.001 A
CH2
0.001 A
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
59