RC Commands
R&S
®
OSP
269
User Manual 1178.7100.02 ─ 05
Example:
Setting
:
CONF:MAIN:TEXT "EMC Test Rack 2"
Writes the text
EMC Test Rack 2
into the "Customer Information"
field.
Example:
Setting
:
CONF:MAIN:TEXT ""
Removes the "Customer Information" field from the "Main" page.
Example:
Query
:
CONF:MAIN:TEXT?
Return string:
"EMC Test Rack 2"
Returns the content of the "Customer Info" field.
Manual operation:
See
CONFigure:RELay:DELay
<channel list>
Sets or queries the command delay times for up to 4 external power-transfer relays
connected to the
OSP-B104 Digital I/O Module (EMS)
period of time, which is reserved for a relay to change its state. Note that these exter-
nal relays require switching times that are significantly longer than in most other relays.
After receiving a
command for changing the state of a connected exter-
nal relay, the module R&S
OSP-B104 behaves as follows:
●
It sends the switching pulse during this full period of time
●
Then it queries the relay's current position
●
If the current position differs from the target position, the module generates a SCPI
error, which is available via
●
Then it accepts the next command
You can set delay times to ensure that the switching process of external transfer relays
is completed, before further commands are executed. If you modify a delay time, the
new value is stored durable on the module's EEPROM memory.
CONFigure Commands