Remote Control Commands
R&S
®
NGL200/NGM200
131
User Manual 1178.8736.02 ─ 09
// ************************************************
// Unlink linked overcurrent protection
// ************************************************
// queries the combined overcurrent protection of the selected channel
CURR:PROT:UNLink 2
CURR:PROT:LINK?2
[SOURce:]CURRent:PROTection:CLEar
...........................................................................131
[SOURce:]CURRent:PROTection:DELay:INITial
................................................................ 131
[SOURce:]CURRent:PROTection:DELay
.......................................................................... 132
[SOURce:]CURRent:PROTection:LINK
.............................................................................132
[SOURce:]CURRent:PROTection:TRIPped?
..................................................................... 133
[SOURce:]CURRent:PROTection:UNLink
......................................................................... 133
[SOURce:]CURRent:PROTection[:STATe]
.........................................................................133
................................................................................................ 134
[SOURce:]CURRent:PROTection:CLEar
[<Channel list>]
Resets the OCP state of the selected channel. If an OCP event has occurred before,
the reset also erases the message on the display.
Setting parameters:
<Channel list>
<list>
Example:
CURR:PROT:CLE (@1)
Resets OCP state at channel 1.
Example:
See
Example "Configuring the overcurrent protection"
Usage:
Setting only
[SOURce:]CURRent:PROTection:DELay:INITial
<duration>
[SOURce:]CURRent:PROTection:DELay:INITial?
[<Channel list>]
Sets or queries the initial fuse delay time once output turns on.
Parameters:
<duration>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for initial fuse delay.
MIN | MINimum
Min value for initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range:
0.00 to 60.00
Configuration Commands