21
PARAMETER
LIST
SETUP DEFAULT
DESCRIPTION
NUMBER
--------------------------------------------------------------------------
SEtuP 01
Setup Parameter Number
Jumps directly to desired parameter by number(1->27).
01
CALSEt F
Scale Setup Configuration and Calibration
Allows for Scale configuration and initial
calibration. This procedure is used when first
connecting a new scale platform to the XC780. The
number of scales in the system, the scale capacities,
and the incremental weight(graduation size) of each
scale can be configured. Select:
F(alse) - If not configuring scale(s) or doing
initial calibration.
t(rue) - If configuring scale(s) or doing initial
calibration.
See the "Calibration" section before proceeding with
calibration.
02
SCLCAL F
Quick Scale Calibration and Fine Span Adjust
Allows for quick calibration and fine adjustment of
span. Quick calibration is used to re-calibrate a
previously calibrated scale during service
maintenance or routine checkup. Fine span adjustment
may be used to correct small span errors(due to load
cell drift or calibration errors) without performing
a complete re-calibration. Most calibration can be
performed using quick calibration. Select:
F - If not calibrating or adjusting scale.
t - If doing quick calibration or adjusting span.
See the "Calibration" section before proceeding with
calibration.
03
PCt 0.0
Minimum PerCent Sample loading
Selects the minimum weight, in percentage of sample
scale capacity, required for accurate sampling(the
sample weight required to turn of the add message).
Select:
0.0 - If minimum is 1 display graduation.
0.2 - If minimum is 0.2% of full scale capacity
0.5 - If minimum is 0.5% of full scale capacity
1.0 - If minimum is 1.0% of full scale capacity
04
trAC F
Automatic Zero Tracking
Selects whether the XC780 will automatically zero the