CALIBRATION PROCEDURE

NI PXIe-4353

This document contains information about verifying and adjusting National Instruments NI PXIe-4353 
modules using NI-DAQmx 9.3 and later. It also contains information about verifying systems that 
include both the NI PXIe-4353 and TB-4353. For more information about calibration, visit 

ni.com/calibration

.

Contents

Software ............................................................................................................................................... 1
Documentation..................................................................................................................................... 2
Calibration Interval .............................................................................................................................. 2
Test Equipment .................................................................................................................................... 2
Test Conditions .................................................................................................................................... 3
NI PXIe-4353 Calibration Procedure .................................................................................................. 3

Initial Setup for the NI PXIe-4353 .............................................................................................. 4
Verifying Module Operating Temperature Changes ................................................................... 4
Verification Procedure ................................................................................................................. 5

Analog Input Voltage Accuracy Verification ...................................................................... 5
Analog Input Offset Verification ......................................................................................... 7
CJC Resistance Accuracy Verification................................................................................ 8

Module Calibration Adjustment .................................................................................................. 10

Test Limits ........................................................................................................................................... 13

Module Verification Test Limit Tables ....................................................................................... 13

Where to Go for Support ..................................................................................................................... 15

Software

Install NI-DAQmx 9.3 or later on the calibration computer. NI-DAQmx includes high-level function 
calls to simplify the task of writing software to calibrate modules. You must have the proper module 
driver installed on the calibration system before calibrating the module.

Note

NI recommends that you install the NI-DAQmx driver software before physically installing 

the NI PXIe-4353. NI-DAQmx, available at 

ni.com/downloads

, configures and controls the 

NI PXIe-4353.

NI-DAQmx supports a number of programming languages, including LabVIEW, LabWindows

/CVI

C/C++, C#, and Visual Basic .NET.

You can access the NI-DAQmx header file, 

NIDAQmx.h

, like any standard library. You can find 

examples of how to use the NI-DAQmx driver by entering the Info Code 

NIDAQmxExLoc

 at 

ni.com/info

 there you will find the topic 

Where Can I Find NI-DAQmx Examples?

ni.com/manuals

De

u

tsch

Français

Summary of Contents for PXIe-4353

Page 1: ...ts 13 Module Verification Test Limit Tables 13 Where to Go for Support 15 Software Install NI DAQmx 9 3 or later on the calibration computer NI DAQmx includes high level function calls to simplify the...

Page 2: ...u use the instruments in Table 1 for calibrating an NI PXIe 4353 module Table 1 Recommended Equipment Equipment Recommended Model Requirements Voltage calibrator Fluke 5520A If this instrument is unav...

Page 3: ...re This section provides instructions for verifying the voltage accuracy performance of the NI PXIe 4353 The calibration process consists of the following steps 1 Initial Setup for the NI PXIe 4353 In...

Page 4: ...ration adjustment can cause you to incorrectly calibrate your module The onboard temperature sensor on the module can be used to verify that this condition does not exist Complete the following steps...

Page 5: ...4353 while referring to Figure 1 Figure 1 CAL 4353 to Fluke 5520A Calibrator Connections 2 Open a DAQmx calibration session on your device 3 Call the DAQmx Connect SC Express Calibration Accessory Cha...

Page 6: ...els you want to verify 7 Set the autozero mode to Every Sample and disable open thermocouple detection 8 On the DAQmx Channel property node set the DAQmx ADC timing mode property to Custom and set the...

Page 7: ...e of the analog input offset of the module 1 Open a DAQmx calibration session on your device 2 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the...

Page 8: ...bration session on your device 4 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the connection to _cjc_calibration and the physical channels to t...

Page 9: ...Limit and Lower Limit values in Table 7 If the result is between these values the channel passes the test 13 Clear the acquisition 14 Repeat steps 5 through 12 for all CJC channels and test point valu...

Page 10: ...mx Connect SC Express Calibration Accessory Channels VI Set the connection to _tc_calibration and the physical channels to Dev1 ai0 31 4 Set the calibrator to output 70 mV and wait for the calibrator...

Page 11: ...0 Set the resistance calibrator to 2 wire compensation mode 11 Configure the CAL 4353 using the DAQmx Connect SC Express Calibration Accessory Channels VI Set the connection to _cjc_calibration and th...

Page 12: ...ose External Calibration VI and set the action to Commit This VI also saves the date time and temperature of the adjustment to the onboard memory Note If an error occurs during adjustment no constants...

Page 13: ...its Contains the Upper Limits and Lower Limits for the test point value That is when the module is within its 1 year calibration interval the test point value should fall between these upper and lower...

Page 14: ...02 3 0 000002 0 000002 4 0 000003 0 000003 5 0 000005 0 000005 6 0 0000065 0 0000065 7 High Speed 0 0000075 0 0000075 Table 6 NI PXIe 4353 Offset with Autozero Disabled Verification Test Limits Timing...