If you want to include this burst signal in your GP-IB program use the
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Pulse stream operating mode is required, but as we start from a
# default status it is not necessary to send a command for setting the
# instrument into pulse stream operating mode.
# Set the period to 20 ns, the pulse width to 10 ns, the leading edge to
# 3 ns and the trailing edge to 5 ns.
:PULSe:PERiod 20NS
:PULSe:WIDTh1 10NS
:PULSe:TRANsition1 3NS
:PULSe:TRANsition1:TRAiling:AUTO OFF
:PULSe:TRANsition1:TRAiling 5NS
N
N
N
N O
O
O
O T E
T E
T E
T E
For Agilent 81110A with Agilent 81112A 3.8V/330 MHz output channel
installed the transitions can be selected either 800 ps or 1.6 ns. Leading
and trailing edge are equal.
# Set the amplitude to 3.3 Volts, the offset to 1.65 Volts.
:VOLTage1 3.3V
:VOLTage1:OFFSet 1.65V
# Enable the output 1 and the complement output 1 (Agilent 81110A
# with Agilent 81112A 3.8V/330 MHz output channel installed, only).
:OUTPut1 ON
:OUTPut1:COMPlement ON