AVW1/AVW4

4-2

water.  The water depth above the sensor is
referred to as the "Reading" in the following
equation.  The Reading decreases with
increasing "Distance" from lip of well to water
surface so the Distance is computed by
subtracting the Reading from the Offset as
shown in the above figure and the following
equation.

Distance = Offset - Reading

FIGURE 4.3-1.  Well Monitoring Example

The "Initial Distance" to the water surface is
measured with a chalked line to be 47.23 feet
below the lip.  The "Initial Reading" is 54.832

feet (of head above the sensor).  Solving the
above equation for the Offset, one obtains:

Offset = Initial Di Initial Reading

= 47.23 feet + 54.832 feet

= 102.062 feet

The following is a sample program that
measures the temperature and frequency of the
vibrating wire sensor and stores the
temperature (

o

C), pressure (psi), temperature

corrected pressure (psi), correction factor (psi),
temperature corrected pressure (feet of H2O),
and Distance from the lip to the water surface
(feet) in Input Locations 1-6 respectively.  This
example assumes the sensor has been
connected as shown in Figure 4.2-1.

Program:  AVW & CR10 USED TO MEASURE
1 GEOKON VIBRATING WIRE SENSOR.

*

1

Table 1 Programs

01:

10

Sec. Execution Interval

01:

P4

Excite,Delay,Volt(SE)

01:

1

Rep

02:

15

2500 mV fast Range

03:

1

IN Chan

04:

1

Excite all reps w/EXchan 1

05:

1

Delay (units .01sec)

06: 2500

mV Excitation

07:

1

Loc [:TEMP  C]

08:

.001

Mult

09:

0.0000

Offset

02:

P55

Polynomial

01:

1

Rep

02:

1

X Loc TEMP  C

03:

1

F(X) Loc [:TEMP  C]

04:

-104.78

C0

05:

378.11

C1

06:

-611.59

C2

07:

544.27

C3

08:

-240.91

C4

09:

43.089

C5

Summary of Contents for AVW4

Page 1: ...RUCTION MANUAL Use of the AVW1 and AVW4 with Geokon Model 4500 Vibrating Wire Piezometers and Pressure Transducers Revision 1 92 C o p y r i g h t c 1 9 8 7 1 9 9 2 C a m p b e l l S c i e n t i f i c I n c ...

Page 2: ...warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization The following contact information is for US and International customers residing in countries served by Campbell Scientific Inc directly Aff...

Page 3: ...MEASUREMENT 3 1 General 3 1 3 2 Multiplier and Offset 3 1 3 3 Swept Frequency Start and End 3 2 3 4 Resolution vs of cycles 3 3 3 5 Temperature Correction 3 3 3 6 Delay Between Measurements 3 4 4 THE AVW1 4 1 General 4 1 4 2 Sensor Hook Up 4 1 4 3 Well Monitoring Example 4 1 5 THE AVW4 5 1 General 5 1 5 2 Sensor Hook Up 5 2 5 3 Datalogger Programming 5 2 6 THE AVW1 AND A MULTIPLEXER Not yet availa...

Page 4: ... Measurement Error on a 3000 Foot Lead 2 4 2 2 4 Temperature Measurement Error on a 5000 Foot Lead 2 5 2 2 5 Thermistor Linearization Error 2 5 2 3 1 Direct Measurement of the Geokon Thermistor 2 6 3 1 1 A Vibrating Wire Sensor 3 1 4 1 1 The AVW1 4 1 4 2 1 Hook up to AVW1 4 1 4 3 1 Well Monitoring Example 4 2 5 1 1 The AVW4 5 1 5 2 1 Hook up for AVW4 5 2 A 1 Pressure Conversion Chart A 1 B 1 1 AVW...

Page 5: ...ll Geokon s manual should be consulted for information on sensor selection and installation Geokon s address and phone number are 48 Spencer Street Lebanon NH 03766 603 448 1562 1 1 SENSOR SELECTION The vibrating wire sensors may be purchased as either vented or sealed sensors The vented sensors have a small hollow vent tube that connects the hollow chamber behind the diaphragm to the atmosphere T...

Page 6: ... some cases there is no measured reference In order to obtain the correct offset under these conditions lower the sensor to a point just above the water level and wait 5 minutes Use the multiplier an offset of 0 0 and the temperature correction function determined in Sections 3 2 and 3 5 to obtain a reading Record the reading Solve equation 3 2 3 for the new offset by plugging in the reading in th...

Page 7: ...to 60oC The precision of the bridge resistors 0 1 results in a tolerance of 0 03oC The accuracy of the datalogger s voltage measurement 0 015 results in a tolerance of 0 01oC The temperature coefficient of the bridge resistors 10 ppm oC results in a tolerance of 0 0003oC oC Errors four five and six mentioned above are all less than 0 03oC each and can probably be ignored The wire resistance is pri...

Page 8: ...05785 0 057855 21 3583 1 304393 21 06127 0 061275 22 3426 1 326119 22 06310 0 063105 23 3277 1 347418 23 06048 0 060484 24 3135 1 368363 24 05747 0 057474 25 3000 1 388888 25 05167 0 051679 26 287 1 408926 26 04042 0 040428 27 2750 1 428571 27 02901 0 029014 28 2633 1 447932 28 02396 0 023968 29 2523 1 466619 29 00577 0 005777 30 2417 1 485089 29 99901 0 00098 31 2317 1 502945 30 98300 0 01699 32 ...

Page 9: ... 53 965 1 794687 53 09674 0 096746 54 929 6 1 803855 54 08849 0 088499 55 895 8 1 812697 55 07032 0 070322 56 863 3 1 821281 56 04819 0 048193 57 832 2 1 829571 57 01651 0 016519 58 802 3 1 837613 57 97896 0 02103 59 773 7 1 845372 58 92977 0 07022 60 746 3 1 852867 59 86962 0 13037 FIGURE 2 2 1 Temperature Measurement Error at Three Temperatures as a Function of Lead Length Wire is 22 AWG with 16...

Page 10: ... 4 FIGURE 2 2 2 Temperature Measurement Error on a 1000 foot Lead Wire is 22 AWG with 16 ohms per 1000 feet FIGURE 2 2 3 Temperature Measurement Error on a 3000 foot Lead Wire is 22 AWG with 16 ohms per 1000 feet ...

Page 11: ...AVW1 AVW4 2 5 FIGURE 2 2 4 Temperature Measurement Error on a 5000 foot Lead Wire is 22 AWG with 16 ohms per 1000 feet FIGURE 2 2 5 Thermistor Linearization Error ...

Page 12: ...his manual covers AVW1 s with serial number 1093 and up plus serial numbers 1002 1034 1040A 1041 1042 1051 1052 1055 1057 1058 1059 1069 1071 1073 1076 1080 1084 1086 1087 1088 1088A 1089A 1090A 1091A and 1092A This manual also covers AVW4s with serial number 1045 and up plus serial numbers 1034 and 1038 These serial numbers mark a change in circuitry that changes the excitation voltage for the te...

Page 13: ...ery short time The wire will vibrate with the resonant frequency for a relatively long period of time and as it does so it will cut the lines of flux in the plucking and pickup coils inducing the same frequency on the lines to the CR10 After waiting for the non resonant frequencies to die out 20 ms the CR10 accurately measures how much time it takes to receive a user specified number of cycles Kno...

Page 14: ...und on the Calibration Sheet Example Using sensor number 3998 the multiplier offset and equation for pressure would be M 1000 digits kHz2 0 0151 psi digit M 15 1 psi kHz2 B 0 0 psi 0 0151 psi digit 9431 digits B 142 4 psi P 15 1 psi kHz2 X kHz2 142 4 psi 3 3 SWEPT FREQUENCY START AND END The AVW1 and AVW4 were not designed for use below 1000 Hz In general the starting frequency of the swept freque...

Page 15: ...s timed to determine the period T in ms of the signal being measured The standard deviation is greater when the period is shorter The period is shortest at zero pressure so the Period at zero pressure given in Table 3 2 1 is used for the example below SD 2 15 1psi kHz2 0 00015ms 500 325 6us 1000us ms 3 SD 0 00026246 psi RESOLUTION 0 00078738 psi Resolution improves as the number of cycles measured...

Page 16: ...e required they can be done one of two ways With the AVW4 the repetitions parameter must be set to 2 The instructs the CR10 to excite both sensors delay 20 ms measure the first sensor and then measure the second sensor There is no delay or re application of the excitation between the measurements Back to back measurements require that the sum of the number of cycles measured on each sensor not exc...

Page 17: ...inches high The silk screening for the AVW1 is shown below FIGURE 4 1 1 The AVW1 4 2 SENSOR HOOK UP FIGURE 4 2 1 Hook up to AVW1 4 3 WELL MONITORING EXAMPLE In this example the vibrating wire sensor is used to monitor the Distance from the lip of the well to the water surface in a 150 foot well The water level is expected to stay within 40 to 80 feet of the lip so the 50 psi pressure sensor is pla...

Page 18: ...feet The following is a sample program that measures the temperature and frequency of the vibrating wire sensor and stores the temperature oC pressure psi temperature corrected pressure psi correction factor psi temperature corrected pressure feet of H2O and Distance from the lip to the water surface feet in Input Locations 1 6 respectively This example assumes the sensor has been connected as sho...

Page 19: ... P37 Z X F 01 4 X Loc T To C 02 0698 F Temp Coeff 03 4 Z Loc T To C 06 P33 Z X Y 01 2 X Loc PRESS psi 02 4 Y Loc T To C 03 3 Z Loc Pt psi 07 P37 Z X F Converts psi to negative ft H20 01 3 X Loc Pt psi 02 2 3067 F 03 5 Z Loc Pt FEET H20 08 P34 Z X F Adds the Offset to the negative of the Reading to obtain the Distance from lip of the well to water surface 01 5 X Loc Pt FEET H20 02 102 06 F OFFSET 0...

Page 20: ...AVW1 AVW4 4 4 This is a blank page ...

Page 21: ...g wire sensors temperature and pressure to eight single ended CR10 channels The AVW4 has no quiescent current drain The current drain during the very short 2 4 ms temperature measurement is 4 mA per channel or lower The current drain during each vibrating wire measurement 170 ms to 500 ms is 32 mA The AVW4 is 6 7 inches long by 2 9 inches wide by 1 4 inches high FIGURE 5 1 1 The AVW4 ...

Page 22: ...rometric pressure only and is used to remove atmospheric pressure changes from the readings on the other three sensors This example assumes the sensors have been connected as shown in the sensor hookup diagram Program AVW4 CR10 USED TO MEASURE 4 GEOKON SENSORS THE FIRST SENSOR SERVES AS A BAROMETER AND IS SUBTRACTED FROM THE OTHER THREE TO MEASURE THREE WATER PRESSURES IN PSI EVERY FIVE MINUTES 1 ...

Page 23: ...4 09 0698 B4 06 P87 Beginning of Loop TEMPERATURE CORRECTION 01 00 Delay 02 4 Loop Count 07 P34 Z X F ASSUMES CALIBRATION TEMP 24oC 01 1 X Loc TEMP C 1 02 24 F 03 17 Z Loc T To C 1 08 P36 Z X Y 01 13 X Loc Cpsi C 1 02 17 Y Loc T To C 1 03 17 Z Loc T To C 1 09 P33 Z X Y 01 5 X Loc psi 1 02 17 Y Loc T To C 1 03 9 Z Loc Pt psi 1 10 P95 End FOLLOWING THREE INSTRUCTIONS SUBTRACT THE BAROMETRIC PRESSURE...

Page 24: ...AVW1 AVW4 5 4 This is a blank page ...

Page 25: ...A 1 APPENDIX A PRESSURE CONVERSION CHART ...

Page 26: ...This is a blank page ...

Page 27: ...B 1 APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 ...

Page 28: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 2 ...

Page 29: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 3 ...

Page 30: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 4 ...

Page 31: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 5 ...

Page 32: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 6 This is a blank page ...

Page 33: ... How many times do we have to change the half period by 1 clock cycle to cover the frequency range 1 F1 2 1 2400 2 208333 ms half period 1 F2 2 1 3100 2 161290 ms half period Change in half period 208333 161290 047043 ms of clock cycle increments to cover the 047043 ms half period range is 047043 ms 0008138 ms 57 81 freq increments How much time is there between frequency increments to cover the f...

Page 34: ...This is a blank page ...

Page 35: ...This is a blank page ...

Page 36: ... campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Te...

Reviews: