background image

UM-0086-A2

 

CANgate User’s Manual 

Page 35 

Requested CAN data are not returned 

 Use 

the 

STATUS

 command to check the status of the slot, and double check the slot definition parameters, as 

described above. 

 

Have you actually polled the slot? A common mistake is to simply enter a 

RQST

 command, which by itself will do 

nothing (unless the sample rate parameter is specified). Normally an 

RP

 command is required, which (for slot 0 

definitions) can be appended to the end of the request command eg. 
    

RQSTJ 2 61444; RP

 

 

Try switching on the diagnostic message display using 

DIAG 3

, then poll the request slot. You should see a 

CAN

x

 

TX> 

message. Check whether there is any reply shown from the ECU – it is possible that the ECU does not 

implement this request. 

 

For J1939, it may be necessary to set the CANgate's network address (default is 0) to a value that is unique on the 
network. Use the 

SETADDR

 command for this. Note that CANgate does not implement the network address 

management protocol described in SAE J1939/81. 

GPS data are not returned 

 Use 

the 

STATUS

 command to check the status of the slot, and double check the slot definition parameters, as 

described above. 

 

Ensure that the GPS port baud rate (set using the 

GPSBAUD

 command) matches that of the GPS unit. 

 Use 

SNOOP GPS

 to check the NMEA-0183 strings being sent by the GPS. If you don't see the string you want then 

you may need to send a command to the GPS (using 

GPSSEND

) to enable it. Verify that the required data item within 

the string is present in the field number specified in the slot definition. 

Data Logger reports timeouts or scan errors 

In this case CANgate is connected to the serial sensor port on a DT80 series data logger. 

 

Check that the logger's serial sensor baud rate has been set to match that of CANgate, eg. 

PS=57600

 

 

Check flow control settings – by default, CANgate uses hardware flow control, so the logger serial sensor port should 
also be set to hardware flow control, using 

PS=HWFC

. If CANgate has been configured to use software flow control 

then the logger must be set likewise (

PS=SWFC

), otherwise the XON and XOFF characters may appear in the 

received data stream, which will probably disrupt the parsing of the returned data. 

 

With the logger in pass-through mode, verify that CANgate is configured for normal operation with no extraneous 
output (ie. 

VERBOSE OFF;DIAG 0

). 

 

With the logger in pass-through mode, verify slot definitions using 

STATUS

. Manually send some of the commands 

that the logger is sending in its 

1SERIAL

 channel definitions and check that CANgate is returning data. Check that 

the format of the data matches that expected by the input conversion in the serial sensor control string (eg. if CANgate 
is returning a floating point value then the serial sensor string should be set up to read a floating point value (

%f

) and 

not, for example, an integer (

%d

)) 

 

Alternatively, enable diagnostic mode on the logger (

P56=1

), which will display data received on the serial sensor 

port in conjunction with the input conversions which are endeavouring to parse it. By examining these, it should be 
possible to determine why the input conversions are not working. 

CANgate repeatedly resets 

If CANgate detects a serious firmware problem then it will trigger a hardware reset. If this does not clear the problem then 
repeated resets may occur. If this is the case then to regain control it may be necessary to set the internal DIP switches to the 
Factory Defaults setting (see 

Configuration

 

(P11)

), and cycle the power. 

Error Messages 

If 

Verbose mode

 is selected, CANgate will return a descriptive message in the event that a problem is detected with any 

received command or operational situation. If verbose mode is switched off, no unsolicited messages will be returned; 
erroneous commands will be silently ignored. 

Syntax Errors 

If there is a problem with a command or parameter, a message is returned indicating where the error occurred. For example: 
 

Error: [ SWOOPJ<err> 2 5000 ] 

In this case the problem is clear: a misspelt command; note that the 

<err>

 tag immediately follows the problematic 

command or parameter. 
 

Error: [ 3 RECVJ<err> 2 61444 ] 

Here the command name is OK but it is not valid in this context (eg. a slot number has been specified but CANgate is not in 
Program Mode, ie. 

BEGIN

 has not been seen.) 

 

Error: [ SNOOPJ <err> ] 

In this case the 

<err>

 tag is not attached to any particular parameter – so the problem here is a missing parameter 

(

CANport

). 

Summary of Contents for CANgate

Page 1: ...Intelligent Data Logging Products CANgate User s Manual A guide to programming wiring communications www datataker com...

Page 2: ...EPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT THERMO FISHER DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED ORAL OR WRITTEN WITH RESPECT TO THE INSTRUMENTS INCLUDING WITHOUT LIMITATIO...

Page 3: ...Summary 15 Command Reference 16 Commands and Parameters 16 Slot Definition Commands 16 RECV Receive Standard ID CAN Messages 16 RECVE Receive Extended ID CAN Messages 17 SEND Transmit Standard ID CAN...

Page 4: ...minal Control 30 KWP2000 OBD II ISO 14230 Requests 30 Reading Fault Codes 31 OBD II 31 J1939 31 Using CANgate with a DT8x Data Logger 32 Escaping Control Characters 32 Serial Sensor Direct Mode 33 Usi...

Page 5: ...N ports are hot pluggable and all CANgate settings are saved in non volatile flash memory so that the unit will resume its configured tasks following a power interruption Typical CANgate application A...

Page 6: ...Unit GPS Global Positioning System ID Identifier ISO International Standards Organisation NMEA National Marine Electronics Association OBD On Board Diagnostics PGN Parameter Group Number PID Parameter...

Page 7: ...ows Pin Signal Function 1 5V out 5Vdc output 200mA max 2 GND Ground 3 CAN1 HI CAN port 1 high 4 GPS RXD GPS port Receive Data input 5 CAN1 LO CAN port 1 low 6 CAN2 HI CAN port 2 high 7 CAN2 LO CAN por...

Page 8: ...agram Note By default the CANgate host port operates at a high baud rate 57600 baud At this speed the maximum cable length for the host computer or data logger connection is approximately 5 metres ass...

Page 9: ...ysical layer as defined in ISO 11898 2 SAE J2284 This is by far the most widely used physical layer The important characteristics of this standard are two wire 5V differential signalling bit rate 10kb...

Page 10: ...then an external termination resistor may be required Direct connection to a single ECU In this case a short 1m cable is used so the two 120 resistors are combined into one 60 resistor If a long cabl...

Page 11: ...ff off off x x x x off 38400 off off ON x x x x off 19200 off ON off x x x x off 9600 off ON ON x x x x off Host port baud rate 115200 ON off off x x x x off hardware x x x off x x x off Host port flo...

Page 12: ...sive and active For RECV RECVE RECVJ and GPS slots CANgate passively receives all matching messages then returns the last known data value either at a fixed rate or when requested by the host data log...

Page 13: ...ing Commands CANgate memory slots and other settings are configured by sending textual commands via the host port The available commands are detailed in the Command Reference P16 If CANgate is directl...

Page 14: ...cters after the apostrophe up until the end of the line will be ignored Run Mode and Program Mode CANgate has two operational modes Run Mode is the normal mode of operation Following power up CANgate...

Page 15: ...ddr sampleRate FORMAT options slot RECVJ CANport PGN startByte bit endByte bit ECUaddr priority sampleRate FORMAT options slot RQSTJ CANport PGN startByte bit endByte bit ECUaddr priority sampleRate F...

Page 16: ...ot specified then 8 is assumed If bit is not specified then 1 is assumed sampleRate is the rate at which to return values to the host integer in ms must be multiple of 100ms May also be ALL in which c...

Page 17: ...ed connect the two CAN ports together and set up the RECV slot on the other port Examples SEND 2 0x302 1122FF07 RP This will immediately send a CAN message on CAN port 2 The message s identifier will...

Page 18: ...negative response and indicates that an invalid request was made All messages are transmitted using the ISO 15765 network layer protocol This protocol fragments long messages into multiple CAN frames...

Page 19: ...slot That is it listens for CAN messages with a particular extended identifier extracts the required data field then returns it to the host when polled or at regular intervals or after every message...

Page 20: ...ter D or M or N default N If N normal is specified then the indicated field is returned as is Otherwise the indicated field and the next field are interpreted as a latitude or longitude value with for...

Page 21: ...specifies the data type integer floating point etc and possibly some other formatting details If no conversion specifier is present which is the default the data value will be output in raw hexadecim...

Page 22: ...that the last matching CAN message had the data 01234567AABBCCDD RECV 1 0x100 1 2 returns 0123 CR LF RECV 1 0x100 1 2 FORMAT 100 returns 29100 00 CR LF 0123 hex scaled by 100 RECV 1 0x100 1 2 FORMAT r...

Page 23: ...of responses may not match the order of the memory slots Examples RP 1 150 Poll all defined memory slots RQSTJ 2 65226 3 6 FORMAT f n RP SEND 1 0x142 112233 RP The RP command is often appended when d...

Page 24: ...time all incoming CAN messages on the specified port are received For each distinct CAN identifier that was received the following information is displayed identifier type standard 11 bit STD or exte...

Page 25: ...cified both CAN ports are measured This command measures the traffic load on one or both CAN ports and reports it as a percentage of the theoretical maximum for the bit rate in use It should be noted...

Page 26: ...es 2 8 end Flags c0820000 Sample 0 ms 1 RECV CAN1 RxID 0x0001 RxBytes 1 8 1 1 Flags c0020000 Sample 0 ms 2 RECV CAN1 RxID 0x0002 RxBytes 1 8 1 1 Flags c0020000 Sample 0 ms 3 RECV CAN1 RxID 0x0003 RxBy...

Page 27: ...s sent by CANgate will be displayed eg CAN2 TX 18EC00FF 132C0007 FFEBFE00 which shows that a message has been sent on CAN2 with identifier 0x18EC00FF and the indicated data value if bit 1 1 then all C...

Page 28: ...adcast PGNs J1939 based systems typically have a multitude of SPNs Suspect Parameter Numbers each representing an individual measured quantity or status A group of related SPNs will be broadcast over...

Page 29: ...d causes the memory slot to be immediately polled which will cause the request to be immediately sent Note that it doesn t make sense to poll a RECVJ slot immediately after defining it because at that...

Page 30: ...at any one time Terminal Control The following example is intended to be used with an ANSI VT100 terminal or emulator Two data values are extracted from received CAN messages with ID 0x212 and one fro...

Page 31: ...eg RQST 1 0101 RP 81066060 The most significant 8 bits 81 of the 32 bit result contains the most pertinent information In this case b31 is set which indicates that the Malfunction Indicator Light MIL...

Page 32: ...ed memory slots a number of 1SERIAL commands within one or more logger schedules typically one 1SERIAL command per parameter of interest These commands will normally send a poll command RP memslot the...

Page 33: ...logger connection from now on will be ignored by the logger and will instead be forwarded to the serial sensor port ie to CANgate Furthermore all normal text output from the logger eg real time data...

Page 34: ...also flash during the snoop If the CAN LED gives a brief flash at two second intervals then this may indicate that bus errors are occurring causing CANgate to disconnect from the bus then retry every...

Page 35: ...pear in the received data stream which will probably disrupt the parsing of the returned data With the logger in pass through mode verify that CANgate is configured for normal operation with no extran...

Page 36: ...message returned if in verbose mode CAN1 BUS ERROR DISCONNECTED The most common cause for this is an incorrectly configured bit rate although it may also be caused by noise or an intermittent electri...

Page 37: ...version 6 Type RESET to clear any existing CANgate configuration 7 Type VXVX All CANgate LEDs will turn on to indicate that the bootloader is active Some random characters may be returned as the bootl...

Page 38: ...1 stop bit Flow Control Not supported Protocols NMEA 0183 Host Data Logger Computer Interface 1 RS232 port Port Speeds 9600 19200 38400 57600 default 115200 baud Serial Parameters No parity 8 data bi...

Page 39: ...r Consumption Idle 0 75W 50mA 15V Maximum CANgate only 2 25W 150mA 15V Maximum CANgate external device drawing 200mA from 5V output 3 75W 250mA 15V Physical Construction Anodised aluminium plastic sur...

Page 40: ...ly to a mode 0x01 request The second byte is the PID and the third and subsequent bytes are the actual data value PID hex Data Size bytes Description Scale Offset Units 00 4 PIDs supported b31 0 PIDs...

Page 41: ...e travelled since DTCs cleared 1 0 km 32 2 Evaporative system vapour pressure 0 25 8192 Pa 33 1 Barometric pressure 1 0 kPa abs bytes 1 2 Sensor 1 WR lambda 1 Equivalence ratio 0 0000305 0 34 4 bytes...

Page 42: ...rain B body C chassis or U network For example P0118 Engine Coolant Temperature Circuit High Input P0203 Injector Circuit Malfunction Cylinder 3 U0003 High Speed CAN Communication Bus Open When a DTC...

Reviews: