Appendix
8.1
Adjustable Functions and Parameters
Setting saved with *SAV n:
Expla-
nation
in
chap.
on
page
Manual setting or via interface
n=1
...
10
n=11
...
255
Setting Command
Significance / Effect
Manu
al
IEEE 488
RS 232C
Default setting
after RESET
*RST:
Function-Specific Commands and Device Settings
Addr
n
Set device address for RS 232 or IEEE 488 (interface configuration)
X
unchanged
bAUd
txt
Set transmission speed (interface configuration)
X
unchanged
txt:
50; 75; 150; ... ; 4800; 9600; 19.2t (transmission speed in bits per sec)
dbit
txt
Set number of data bits (interface configuration)
X
unchanged
txt:
7; 8
DE
LAY
v
Output off delay for OCP
X
X
00.00
v:
0; 0.01; ... , 99.99 (delay time in seconds)
X
X
DI
SPLAY
txt
Activate / deactivate the 7-segment displays
X
X
ON
txt:
ON, OFF
(display on, display off = shorter measuring and setting rate via the interface)
IL
IM
v
Current setting limit (soft-limit)
X
X
I
nom
v:
12.00
(example: 12.00 A)
X
X
IS
ET
v
Current setpoint
X
X
X 000.000
v:
10.75
(example: 10.75 A)
X
X
OC
P
txt
Response selection for current limiting (overcurrent protection)
X
X
OFF
txt:
OFF
(limiting by means of current regulation)
X
X
ON
(limiting by deactivating the output after DELAY time elapses)
OU
TPUT
txt
Output On/Off
X
X
OFF
txt:
OFF, ON
(output deactivated, output activated)
X
X
OVS
ET
v
Overvoltage protection trigger value
X
X
OVSETmax
v:
32.5
(example: 32.5 V)
X
X
Pbit
txt
Set parity bit (interface configuration)
X
unchanged
txt:
nonE
no parity bit
ZEro
always 0
EVEn
even numbered
odd
odd numbered
onE
always 1
Pon txt
Device setting selection after power-up
X
unchanged
POW
ER_ON
txt
txt:
RST
(default setting)
X
X
RCL
(last used settings)
SBY
(last used setting, output deactivated)
rEP n
Set number of repetitions for the SEQUENCE function
X
X
unchanged
RE
PETITION
n
n:
0
(continuous repetition)
X
X
1; 2; ... , 255 (number of sequence repetitions)
rnd
txt
Round off the displayed measured value (setup)
X
unchanged
txt:
0; –1; –2
(no rounding, round by one or two decimal places)
Sbit
txt
Set number of stop bits (interface configuration)
X
unchanged
txt:
1; 2
SEq
txt
Sequence, automatic memory recall
X
running sequence
STOP
SEQ
UENCE
txt
txt:
Go (start); (off, on (close / end, open))
X
X
hold (pause/suspend); cont (resume); strt, step, stop (step-by-step control)
Strt m
Start and stop memory address for the SEQUENCE function
X
X
unchanged
StoP n
m:
11; 12; ... , 255 (start address = beginning of sequence)
STA
RT_STOP
m, n
n:
11; 12; ... , 255 (stop address = end of sequence,
n
≥
m
)
X
X
STO
RE
n, v1, v2, v3, txt
Direct transfer of parameter data to memory
X
X
no effect
n:
11; 12; ... ; 255 (memory address)
v1:
voltage setpoint
v2:
current setpoint
v3:
memory location-specific dwell time
txt:
NC: no function, CLR: complete deletion of memory location data
TD
EF
v
Set default time for the SEQUENCE function
X
X
unchanged
(TDEF is used instead of TSET if TSET has not been set to any specific value, or if it has been set to 0.)
X
X
v:
0.01; 0.02; ... , 99.99 (delay time in seconds)