GDM-8246 MULTIMETER
PROGRAMMER MANUAL
⎯
24
⎯
*SRE?
Returns:
<NR1> is in the range from 0 through 255.
Examples
*SRE 7 sets bits of the SRER to 0000 0111.
If the *SRE? returns 3, the SRER is set to 0000 0011.
*STB?
(query only)
Function:
The query of the Status Byte register (SBR) with *STB? will return a
decimal number representing the bits that are set (true) in the status
register.
Syntax:
*STB?
Returns:
<NR1> is in the range from 0 through 255.
Examples:
*STB? returns 81, if SBR contains the binary value 0101 0001.
*WAI
(no query form)
Function:
WAI prevents the programming instrument from executing further
commands or queries until all pending operations are finished.
Syntax:
*WAI
GDM-8246 MULTIMETER
PROGRAMMER MANUAL
⎯
25
⎯
:CONFigure:AUTo
Function:
Set or query the auto-range mode on or off.
Syntax:
:CONFigure:AUTo?
:CONFigure:AUTo <Boolean>
<Boolean> can be 0(off) or 1(on)
Returns:
0/1
Examples:
:CONFigure:AUTo 1
Set to auto-range mode.
If it is in the auto-range mode, the command of :CONFigure:AUTo? Will
return the value of 1.
:CONFigure:RANGe?(query only)
Function:
Set or query the range of the present function.
Syntax:
:CONFigure:RANGe?
Returns:
<NR2>