PEM200 User Manual
39
Command
:DETector:INTegration #
Short
command
:DET:INT #
Type
Get/Set
Purpose
Set or get the integration time in seconds. The range is 1e-3 (1ms) to 1e2
(10s).
Input example
:DET:INT 0.25
Output
example
[ITIME](2.1)
Command
:DETector:Measure?
Short
command
:DET:M?
Type
Get
Purpose
Starts the synchronous detection system and returns the magnitude of the
harmonic components as set by :DET:FORM
Input example
:DET:M
Output
example
[MXY](1.314652E0, 3.623624E-1,2.1465E-1,1.02436E-4,-9.21565E-5)
Command
:DETector:MeasureDC?
Short
command
:DET:MDC?
Type
Get
Purpose
Quickly measure the DC amplitude of the input optical signal. Returns mean
and standard deviation.
Input example
:DET:MDC?
Output
example
[DC](1.314652E2, 0.244652E0)