6.2 ET command set (Compatibility to old HPS)
This command set is selected via Menu “F10 Set Instruct” “ET”.
Setting commands
Command:
Command example:
Command to set an output voltage:
U,<Voltage>kV
U,1.000kV
Command to set a voltage limit:
UL,<Voltage>kV
UL,1.000kV
Command to set an output current:
I,<Current>mA
I,30mA
Command to set a current limit:
IL,<Current>mA
IL,30mA
Command to set the Voltage ramp speed:
RAMP,<Rampspeed>V/s
RAMP,3000V/s
Command to switch the high voltage on and off:
HV,ON
HV,OFF
Command for setting the KILL function „Enable“ or „Disable“:
KILL,ENable
KILL,EN
KILL,DISable
Command for Emergency off (The HV generation is shut off permanently and the set values for volt-
age and current are set to Zero):
EMCY OFF
Read out the setting commands
Command:
Response example:
Read out the set voltage:
STATUS,U
U,
RANGE=3.000kV,
VALUE=2.458kV
Read out the voltage limit:
STATUS,UL
UL,
RANGE=3.000kV,
VALUE=2.458kV
Read out the set current:
STATUS,I
I,
RANGE=100.0mA,
VALUE=30.0mA
Read out the current limit:
STATUS,IL
IL,
RANGE=100.0mA,
VALUE=30.0mA
Read out the voltage ramp speed:
STATUS,RAMP
RAMP,
RANGE=3000V/s,
VALUE=1000V/s