background image

 

 

 

               

 

 

14

  

 

 

 

UM200002 

ISSUE

 A 

REV

 00 

User Manual 

7.5.7.4

 

Password 

This screen is used to configure the Password FTP server. 

 

7.5.7.5

 

Remote Dir 

This screen is used to configure the Remote Directory of 
FTP server where files will be uploaded from data-logger. 

 

 

7.6

 

Discharge Table Configuration 

In  order  to  calculate  discharge,  we  must  first  configure 
the Discharge table in LogPRO TOUCH device. Discharge 
table can be easily configured following the steps below: 

1.

 

Create a text file discharge-config.xml and open it 
in any simple text editor like Notepad. 

<discharge-config> 
  <levelTag>Level</levelTag> 
  <surfaceVelocityTag>Velocity 
  </surfaceVelocityTag> 
  <map> 
    <add>0.00,0.00,0.89</add> 
    <add>0.48,1.76,0.89</add> 
    <add>0.48,1.76,0.89</add> 
    <add>0.49,1.82,0.89</add> 
    <add>0.50,1.87,0.89</add> 
    <add>0.51,1.92,0.89</add> 
    <add>0.52,1.97,0.89</add> 
    <add>0.69,2.87,0.89</add> 
    <add>0.70,2.93,0.89</add> 
    <add>0.71,2.98,0.89</add> 
    <add>0.72,3.03,0.89</add> 
    <add>0.75,3.19,0.89</add> 

    <add>0.77,3.30,0.89</add> 
    <add>0.79,3.43,0.89</add> 
    <add>0.80,3.49,0.89</add> 
    <add>0.82,3.62,0.89</add> 
    <add>0.83,3.69,0.89</add> 
    <add>0.84,3.76,0.89</add> 
    <add>0.85,3.84,0.89</add> 
    <add>0.86,3.92,0.89</add> 
    <add>0.87,4.00,0.89</add> 
    <add>0.88,4.07,0.89</add> 
    <add>0.89,4.15,0.89</add> 
  </map> 
</discharge-config> 

 

 
Here: 

Option 

Description 

levelTag 

Name of Tag that measures 
Water Column Level 

surfaceVelocityTag  Name of Tag that measures 

Surface Velocity of Water 

map 

Cross-section and K factor 
mapping for metered water 
column. 
 
NOTE: The table has to be in 
increasing order of water 
column. 
 

<add>0.48,1.76,0.89</add> 
 

 

First value 0.48 
represents the water 
column height (in meters) 

 

Second Value 1.76 
represents the cross-
section area at the water 
column (in m

2

 

Third Value 0.89 
represents the K factor at 
water column 

 

2.

 

Prepare to transfer the file in LogPRO TOUCH. 

a.

 

Arrange  an  empty  USB  pen  drive.  Create 
following folders in USB Pen drive. 
 
AutoCopy > Flashdisk > Firmware > 1.1.0.0 > 
Configs 
 

b.

 

Inside  Config  folder,  save  the  discharge-
config.xml and safely eject USB pen drive. 

3.

 

Apply the new discharge table 

a.

 

Insert  the  USB  pen  drive  in  LogPRO  TOUCH. 
Ensure that device is powered ON.  

b.

 

Wait  for  5  seconds  and  eject  the  USB  pen 
drive. 

c.

 

Restart the LogPRO TOUCH. 

 
 
 

Only simple FTP supported.  
SSL is NOT supported. 
 

 

Summary of Contents for LogPRO TOUCH

Page 1: ...User Manual LogPRO TOUCH Touch based IoT Edge Gateway Data Logger Document Issue Rev Issue Date Prepared By Approved By UM200002 A 00 Aug 2020 MR PROP...

Page 2: ...mation in this document is subject to change without notice and should not be considered a commitment by M2MLogger M2MLogger assumes no responsibility for any errors that may appear in this document T...

Page 3: ...iring Plan for MODBUS RS 485 10 7 4 3 Wiring Plan for Analog Inputs 4 20mA 10 7 5 User interface 11 7 5 1 Home 11 7 5 2 Backup 12 7 5 3 Settings 12 7 5 4 Interval 12 7 5 5 Ethernet 12 7 5 6 GPRS 13 7...

Page 4: ...LOST PROFITS LOST SAVINGS OR INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OR INABILITY TO USE ANY OF THE PRODUCTS EVEN IF M2MLOGGER HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES O...

Page 5: ...w velocity in m s and A is cross section area in m2 For real world measurements it is important to understand that the velocity of the moving water varies both across the stream channel and from the s...

Page 6: ...6 UM200002 ISSUE A REV 00 User Manual 5 Components...

Page 7: ...7 UM200002 ISSUE A REV 00 User Manual 6 Cable Wiring Diagram LOGPRO TOUCH...

Page 8: ...ver The Air OTA upgrades LogPRO supports MODBUS RS 485 through screw terminals USB pen drive through USB type A connector 10 100 Mbps Ethernet through a standard Ethernet connector RJ 45 and GSM GPRS...

Page 9: ...9 UM200002 ISSUE A REV 00 User Manual 7 3 Mechanical Figure 3 Physical Dimensions mm...

Page 10: ...Shielded Ground 7 4 3 Wiring Plan for Analog Inputs 4 20mA Pin Function SP Switched Power Output SP0 SP3 AI Analog Input AI0 AI15 G Shielded Ground AC DC Line Voltage 24VDC GND LOGPRO Power Supply Fi...

Page 11: ...H HSDPA HSUPA TDSCDMA HSPA H HSPA 4G LTE LTE A TDD LTE FDD LTE airtel Carrier 14 4V Battery Voltage Battery Charging Status No Icon Functionality not supported Gray Battery not charging Green Battery...

Page 12: ...interval of device Option Description 1 min record every 1 minute 2 min record every 2 minutes 5 min record every 5 minutes 10 min record every 10 minutes 15 min record every 15 minutes 7 5 5 Etherne...

Page 13: ...sed to configure the APN for Internet access 7 5 6 2 Dialup This screen is used to configure the Dialup number for internet connection 7 5 7 FTP This window shows device FTP settings 7 5 7 1 URL This...

Page 14: ...d 0 82 3 62 0 89 add add 0 83 3 69 0 89 add add 0 84 3 76 0 89 add add 0 85 3 84 0 89 add add 0 86 3 92 0 89 add add 0 87 4 00 0 89 add add 0 88 4 07 0 89 add add 0 89 4 15 0 89 add map discharge conf...

Page 15: ...ully 4 stage PWM charge management Build in short circuit protection open circuit protection reverse protection over load protection Dual MOSFET reverse current protection low heat production 8 3 2 Sp...

Page 16: ...16 UM200002 ISSUE A REV 00 User Manual 8 4 Tripod Dimensions...

Page 17: ......

Page 18: ...1 B Friends Colony Industrial Area Shahdara Delhi 110095 India Telephone 91 11 22112232 WORKS B 8 First Floor FIE Patparganj Industrial Area Delhi 110092 India Telephone 91 11 22582090 email sales m2m...

Reviews: