16
|
ni.com
|
NI 449
x
Calibration Procedure
3.
Initialize the AI calibration using the DAQmx Initialize External Calibration VI, as shown
below.
Note
Refer to the NI-DAQmx function parameters for the LabVIEW input values.
4.
Perform the AI calibration adjustment by selecting the DAQmx Adjust DSA AI Calibration
VI from the DAQmx Adjust DSA AI Calibration Polymorphic VI, as shown below.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxInitExtCal
with the
following parameters:
deviceName
:
Dev_name
password
:
NI
calHandle
:
&myCalHandle
LabVIEW Block Diagram
NI-DAQmx Function Call
In DAQmx 9.1.7 or earlier, the VI will look like
this:
In DAQmx 9.2 or later, the VI will look like this
(as long as the Polymorphic VI selector is
visible):
Call
DAQmxAdjustDSAAICal
with the
following parameters:
calHandle
:
myCalHandle
referenceVoltage
:
0.2