NI 9237 Calibration Procedure
8
ni.com
2.
Set the calibrator voltage to the Calibrator Output value indicated in
Table 4. NI recommends that you verify all values, although you can
save time by verifying only the values used in your application.
3.
If you use C function calls, create a task using
DAQmxCreateTask
, as
shown in the following table. If you use LabVIEW, skip this step. The
task is created in step 4 in LabVIEW.
LabVIEW Block Diagram
NI-DAQmx Function Call
LabVIEW does not require this step.
Call
DAQmxCreateTask
with
the following parameters:
taskName
:
AIVerificationTask
taskHandle
:
&taskHandle