NI 4070/4072 Calibration Procedure
54
ni.com
4.
Call
niDMM_Read
to configure the NI 4070/4072 for a current mode
before applying current.
5.
Output 20 mA on the calibrator with the current output set to
NORM
.
6.
Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
0.02
7.
Output –20 mA on the calibrator with the current output set to
NORM
.
8.
Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
= –
0.02
9.
Output 0 A on the calibrator.
10. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
11. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_AC_CURRENT
•
Range
=
0.01
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
12. Output 200 mA on the calibrator.
13. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
0.2
14. Output –200 mA on the calibrator.
15. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
= –
0.2