NI 5105 Calibration Procedure
|
© National Instruments
|
11
Note
If any of these tests fail immediately after you perform an external
adjustment, make sure that you have met the requirements listed in the
sections before you return the device to
National Instruments for repair.
Vertical Offset Accuracy
Table 5 contains the input parameters for verifying vertical offset accuracy, the calibration test
limits, and the published specifications of the NI 5105.
To verify vertical offset accuracy, complete the procedures described in this section for each of
the 18 iterations listed in Table 5 for channel 0, then repeat the procedure for each of the
remaining 7 channels on the NI 5105.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
Note
Throughout the procedure, refer to the C/C++ function call parameters for the
LabVIEW input values.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE