NI 5105 Calibration Procedure
|
© National Instruments
|
31
2.
Connect the scope calibrator test head directly to the channel input of the device. Configure
the calibrator to output the voltage listed under the
Input (V)
column in Table 9. Configure
the load impedance of the calibrator to 1 M
Ω
.
3.
Wait 2,500 ms for the impedance matching and frequency of the calibrator to settle.
4.
Adjust the vertical range using the niScope Cal Adjust Range VI.
5.
Repeat steps 2 through 4 for each iteration listed in Table 9.
6.
Move the calibrator test head to the next device input channel and repeat steps 2 through 5.
Update the
channelName
parameter for each channel being adjusted.
7.
Connect the calibrator test head directly to the channel 0 input of the device. Configure the
calibrator to output an exact 3.0 MHz sine wave with 1 V
pk-pk
amplitude and 50
Ω
load
impedance.
8.
Adjust the sample rate of the device using the niScope Cal Adjust VCXO VI.
Note
The 3 MHz stimulus is automatically taken from channel 0.
LabVIEW VI
C/C++ Function Call
Call
niScope_CalAdjust
Range
with the following
parameters:
sessionHandle
: The session
handle from
niScope_CalStart
channelName
: The channel
being adjusted
range
: The
Range
value listed
in Table 9 for the current
iteration
stimulus
: The
Input Voltage
value listed in Table 9 for the
current iteration
LabVIEW VI
C/C++ Function Call
Call
niScope_Cal
AdjustVCXO
with the
following parameters:
sessionHandle:
The
session handle from
niScope_CalStart
stimulusFreq
:
3,000,000